耳温计检定校准用黑体空腔有效发射率比较与亮度温度比对实验

杨雪 柏成玉 王景辉 原遵东

杨雪,柏成玉,王景辉,等. 耳温计检定校准用黑体空腔有效发射率比较与亮度温度比对实验[J]. 计量科学与技术,2021, 65(9): 40-43 doi: 10.12338/j.issn.2096-9015.2020.0105
引用本文: 杨雪,柏成玉,王景辉,等. 耳温计检定校准用黑体空腔有效发射率比较与亮度温度比对实验[J]. 计量科学与技术,2021, 65(9): 40-43 doi: 10.12338/j.issn.2096-9015.2020.0105
YANG Xue, BAI Chengyu, WANG Jinghui, YUAN Zundong. Blackbody Cavity Effective Emissivity Comparison and Brightness Temperature Comparison Experiment for Ear Thermometer Verification and Calibration[J]. Metrology Science and Technology, 2021, 65(9): 40-43. doi: 10.12338/j.issn.2096-9015.2020.0105
Citation: YANG Xue, BAI Chengyu, WANG Jinghui, YUAN Zundong. Blackbody Cavity Effective Emissivity Comparison and Brightness Temperature Comparison Experiment for Ear Thermometer Verification and Calibration[J]. Metrology Science and Technology, 2021, 65(9): 40-43. doi: 10.12338/j.issn.2096-9015.2020.0105

耳温计检定校准用黑体空腔有效发射率比较与亮度温度比对实验

doi: 10.12338/j.issn.2096-9015.2020.0105
基金项目: 国家质量基础的共性技术研究与应用国家重点研发计划(2017YFF0205504);首都科技条件平台(科研)(Z181100009518012);国家质检总局科技计划项目(2015QK001)。
详细信息
    作者简介:

    杨雪(1983-),中国计量科学研究院工程师,研究方向:辐射测温,邮箱:yangx@nim.ac.cn

    通讯作者:

    原遵东(1960-),中国计量科学研究院研究员,研究方向:辐射测温,邮箱:yuanzd@nim.ac.cn

Blackbody Cavity Effective Emissivity Comparison and Brightness Temperature Comparison Experiment for Ear Thermometer Verification and Calibration

  • 摘要: 耳温计检定校准用黑体辐射源是耳温计检定和校准的计量标准器,黑体空腔的有效发射率是其关键特性。介绍耳温计检定规程推荐黑体空腔的发射率特性,并与两种国际标准推荐的耳温计黑体空腔进行了亮度温度比对实验。在数据分析基础上,给出对三个耳温计黑体空腔的评价。
  • 图  1  NIM G2空腔

    Figure  1.  Cavity of NIM G2

    图  2  ASTM空腔

    Figure  2.  ASTM Cavity

    图  3  EN空腔

    Figure  3.  EN Cavity

    图  4  耳温计检定校准用黑体空腔的出射角ω

    Figure  4.  The exit angle ω of blackbody cavity for ear thermometer calibration

    图  5  测试模式下耳温计测量重复性

    Figure  5.  Measurement repeatability of ear thermometer in test mode

    表  1  耳温计检定校准用黑体辐射源空腔规格参数

    Table  1.   Cavity parameters of blackbody radiation source for verification and calibration of ear thermometer

    温度
    范围
    开口
    直径
    腔体
    直径
    深度漆发
    射率
    腔口100°出射
    角内发射率
    20 °C~50 °C 10 mm 60 mm 118.2 mm0.9650.9998
    0.940.9997
    0.920.9996
    下载: 导出CSV

    表  2  标准铂电阻温度计示值比较(单位: ℃)

    Table  2.   Comparison of indication of standard platinum resistance thermometer (unit: ℃)

    温度/ ℃一等标准铂电阻
    温度计1# ℃
    一等标准铂电阻
    温度计2#/ ℃
    35.00035.00235.002
    37.00037.00237.003
    41.50041.50042.502
    下载: 导出CSV

    表  3  亮度温度比对结果(单位: ℃)

    Table  3.   Comparison results of repeatability brightness temperature measured by ear thermometer in test mode (unit: ℃)

    温度 亮度温度差1(ASTM-NIM) 亮度温度差2(EN-NIM)
    测试1 测试2 平均值 测试1 测试2 平均值
    35.50 −0.01 0.00 0.00 −0.01 −0.01 −0.01
    37.00 0.00 −0.01 0.00 −0.01 −0.01 −0.01
    41.50 0.00 0.00 0.00 −0.01 −0.01 −0.01
    下载: 导出CSV
  • [1] ASTM. Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature: ASTM Standards E 1965-98[S].West Conshohocken, PA: ASTM, 1998.
    [2] European Committee for Standardization. Clinical Thermometers—Part 5: Performance of Infra-Red Ear Thermometers (with Maximum Device): EN 12470-5:2003[S]. CEN, Brussels, 2003.
    [3] Japanese Industrial Standard Infrared Ear Thermometers: JIS T 4207[S].Tokyo, 2005.
    [4] 国家技术监督局. 红外耳温计检定规程: JJG1164-2019[S]. 北京: 中国质检出版社, 2019.
    [5] 国家技术监督局. 工作用辐射温度计检定规程: JJG856-2015[S]. 北京: 中国质检出版社, 2015.
    [6] Kim G J , Yoo Y S , Kim B H , et al. A small-size transfer blackbody cavity for calibration of infrared ear thermometers[J]. Physiological Measurement, 2014, 35(5):753.
    [7] S F Tsai. Comparison Measurements of Infrared Ear Thermometers Against Three Types of Blackbody Sources[J]. International Journal of Thermophysics,2010,31: 8-9.
    [8] Punik I , Clausen S , Favreau J O , et al. Comparison of Blackbodies for Calibration of Infrared Ear Thermometers[J]. International Journal of Thermophysics, 2011, 32(1-2):127-138.
    [9] 柏成玉, 原遵东, 王铁军. 红外耳温计的校准[J]. 计量技术, 2007(4): 46-49.
    [10] 杨方英, 刘丽华. 红外耳温计的临床应用观察[J]. 护理研究, 2002, 16(1): 8-9.
    [11] 邹轶, 柏成玉, 曾麟, 等. 环温对红外耳温计校准的影响实验[J]. 计量技术, 2019(1): 52-54.
    [12] 周四清, 吴双双, 张莉群, 等. 人体红外温度计校准装置设计与验证[J]. 中国计量, 2016(12): 71-72.