Development of Verification Equipment for Phoropters
-
摘要: 为了解决综合验光仪的计量检定,研制了基于自准直清晰度法的综合验光仪球镜度、柱镜度、柱镜轴位、中心误差光电检测系统。该系统采用光栅测长传感器实现凹面反射镜位移的精密测量,采用角度编码器实现目标十字线分划板角位移的传感,使用CCD实现图像采集,利用计算机实现分析及测量。分析了影响测量精度和范围的因素,设计了综合验光仪的光学系统、关键物镜,给出了光电参数,仪器可溯源至顶焦度国家基准。理论分析和实验结果表明,该装置用于检测综合验光仪时球镜顶焦度扩展不确定度可达到(0.03~0.08) m−1(k=2),满足综合验光仪的检定需求。Abstract: To solve the verification of the phoropters, optoelectronic equipment for measuring spherical vertex power, cylinder vertex power, axis position, and the center error of refractor heads based on the self-collimation sharpness method was developed. The equipment uses a grating length sensor to achieve precision measurement of concave reflector displacement, an angle encoder to measure the angular of the cross-line reticle, a CCD to achieve image acquisition, and a computer to achieve analysis and measurement. In this paper, the factors affecting the measurement accuracy and range are analyzed, the optical system of the comprehensive optometer and the key objective lens is designed, the photoelectric parameters are given, and the equipment is traceable to the national primary standard of vertex power. Theoretical analysis and experimental results indicate that the expanded uncertainly for phoropters can reach (0.03~0.08) m-1(k=2), which meets the verification demand for the phoropters.
-
Key words:
- focusing error /
- back-vertex-power (BVP) /
- prism power /
- verification /
- uncertainty /
- phoropters
-
表 1 测量结果的示值误差
Table 1. Indication error of the measurement results
m−1 标准值 2.49 7.02 12.00 15.96 20.01 测量值 2.48 7.03 11.98 15.97 19.99 示值偏差 −0.01 0.01 −0.02 0.01 −0.02 标准值 −2.52 −6.97 −12.04 −15.98 −20.05 测量值 −2.52 −6.98 −12.03 −15.98 −20.02 示值偏差 0.00 −0.01 0.01 0.00 0.03 表 2 测量重复性引入的标准不确定度
${u_1}\left( M \right)$ Table 2. Uncertainty introduced by measurement repeatability
m−1 球镜顶焦度 标准不确定度 [−3~3] 0.009 (−3~−6]或(3~6] 0.010 (−6~−9]或(6~9] 0.016 (−9~−12]或(9~12] 0.019 (−12~−15]或(12~15] 0.025 (−15~−20]或(15~20] 0.036 表 3 合成标准不确定度
Table 3. Combined standard uncertainty
m−1 球镜顶焦度 扩展不确定度(k=2) [−3~3] 0.0136 (−3~−6]或(3~6] 0.0142 (−6~−9]或(6~9] 0.0190 (−9~−12]或(9~12] 0.0243 (−12~−15]或(12~15] 0.0292 (−15~−20]或(15~20] 0.0390 表 4 测量结果的扩展不确定度
Table 4. The expanded uncertainly of measurement result
m−1 球镜顶焦度 扩展不确定度(k=2) [−3~3] 0.03 (−3~−6]或(3~6] 0.03 (−6~−9]或(6~9] 0.04 (−9~−12]或(9~12] 0.05 (−12~−15]或(12~15] 0.06 (−15~−20]或(15~20] 0.08 -
[1] 郑定列. 自动验光仪—科学验光的发展趋势[J]. 中国眼镜科学杂志, 2008(3): 79-81. [2] 史亚楠. 综合验光仪的应用及注意事项[J]. 计量与测试技术, 2020, 47(9): 65-66. [3] 闫洁. 综合验光仪光学系统研究[J]. 光学仪器, 2010, 32(5): 39-43. [4] 李飞, 李林, 洪宝玉, 等. 基于激光差动共焦技术的后顶焦度测量系统[J]. 计量学报, 2014, 35(2): 129-132. doi: 10.3969/j.issn.1000-1158.2014.02.07 [5] ZHANG J Y, LIU W L, SUN J. Theoretical considerations and measurements for phoropters[C]. Seventh International Symposium on Instrumentation and Control Technology. Beijing: Proc. SPIE, 2008: 712903-1-6. [6] ZHANG J Y, SUN J. GAO M L, et al. Metrological research on test method for phoropters[C]. 8th international conference on electronic measurement and instruments. Beijing: IEEE Express, 2007: 16-18. [7] International Organization for Standardization. Ophthalmic instruments Refractor heads: ISO 10341[S]. Geneva: International Organization for Standardization, 1997. [8] International Organization for Standardization. Optics and optical instruments-Test lenses for calibration of focimeters Part 1: Test lenses for focimeters used for measuring spectacle lenses: ISO 9342-1[S]. Geneva: International Organization for Standardization, 2005. [9] 国家质量监督检验检疫总局. 综合验光仪(含视力表) 检定规程: JJG 1097-2014[S]. 北京: 中国质检出版社, 2014. [10] 朱林泉, 潘德恒. 焦度计的设计和测量中的精度分析[J]. 兵工学报, 1996, 17(3): 266-269. [11] 杨并上, 廖海洋. 眼镜度数全自动测量方法[J]. 重庆大学学报(自然科学版), 2006, 29(10): 65-68. [12] CM Sullivan, CW Fowler. Grating visual acuity testing as a means of psychophysical assessment of progressive addition lenses[J]. Optometry and Vision Science, 1989, 66(9): 565-572. doi: 10.1097/00006324-198909000-00001 [13] Rosenblum W M, O'Leary D K, Blaker W J. Computerized Moiré analysis of progressive addition lenses[J]. Optometry and Vision Science, 1992, 69(12): 936-940. doi: 10.1097/00006324-199212000-00005 [14] 谭宇. 平面光学元件最小焦距的双星点定焦测量法的精度分析[J]. 应用光学, 2001, 22(1): 43-45. [15] 北京工业学院光测组. 双星点定焦法的讨论[J]. 光学工艺, 1975(5): 15-24.