临界流文丘里喷嘴法气体流量标准装置期间核查方法研究

王雪丽 赵作广 徐焰栋 徐昭

王雪丽,赵作广,徐焰栋,等. 临界流文丘里喷嘴法气体流量标准装置期间核查方法研究[J]. 计量科学与技术,2023, 67(1): 68-72 doi: 10.12338/j.issn.2096-9015.2022.0062
引用本文: 王雪丽,赵作广,徐焰栋,等. 临界流文丘里喷嘴法气体流量标准装置期间核查方法研究[J]. 计量科学与技术,2023, 67(1): 68-72 doi: 10.12338/j.issn.2096-9015.2022.0062
WANG Xueli, ZHAO Zuoguang, XU Yandong, XU Zhao. Research on Intermediate Check Method for Gas Flow Standard Devices Using Critical Flow Venturi Nozzles[J]. Metrology Science and Technology, 2023, 67(1): 68-72. doi: 10.12338/j.issn.2096-9015.2022.0062
Citation: WANG Xueli, ZHAO Zuoguang, XU Yandong, XU Zhao. Research on Intermediate Check Method for Gas Flow Standard Devices Using Critical Flow Venturi Nozzles[J]. Metrology Science and Technology, 2023, 67(1): 68-72. doi: 10.12338/j.issn.2096-9015.2022.0062

临界流文丘里喷嘴法气体流量标准装置期间核查方法研究

doi: 10.12338/j.issn.2096-9015.2022.0062
详细信息
    作者简介:

    王雪丽(1989-),江苏省质量技术监督气体流量计量检测中心工程师,研究方向:力学计量、质量管理等,邮箱:wxl_hnu@163.com

Research on Intermediate Check Method for Gas Flow Standard Devices Using Critical Flow Venturi Nozzles

  • 摘要: 期间核查是实验室量值溯源计划体系的重要部分。常用的临界流文丘里喷嘴法气体流量标准装置的期间核查方法是En值判定法,但该方法对核查标准的长期稳定性依赖性很大,容易产生误判。设计了一种装置内喷嘴(组)比对核查与实验室内多套装置比对核查相结合的方法,并以一套临界流文丘里喷嘴法气体流量标准装置为例,进行了核查方法设计和试验数据分析。该方法不但能对装置中各主要计量器具进行全面核查,还能随时对实验室内多套同类装置进行核查,减少核查标准由于安装以及使用环境变化造成的测量偏差,使得临界流文丘里喷嘴法气体流量标准装置的期间核查工作更加便捷、可靠。
  • 图  1  临界流文丘里喷嘴法气体流量标准装置原理图

    Figure  1.  Schematic diagram of critical flow venturi nozzle gas flow standard device

    表  1  装置具体信息

    Table  1.   Information of the gas flow standard devices

    名称测量范围Urelk=2)喷嘴数量
    装置11~20000 m3/h
    DN 50~600 mm
    0.30%29
    装置20.2~5000 m3/h
    DN 15~200 mm
    0.30%18
    装置30.25~1600 m3/h
    DN 15~150 mm
    0.27%14
    下载: 导出CSV

    表  2  装置1临界流文丘里喷嘴配置情况

    Table  2.   The configuration of critical flow venturi nozzles in device 1

    喷嘴编号名义流量(m3/h)喷嘴编号名义流量(m3/h)
    11161024
    22171024
    34181024
    48191024
    516201024
    632211024
    764221024
    8128231024
    9256241024
    10512251024
    111024261024
    121024271250
    131024281250
    141024291250
    151024//
    下载: 导出CSV

    表  3  比对流量点及喷嘴组合方式

    Table  3.   Comparison of flowrate and nozzle combinations

    喷嘴(组)名义流量
    (m3/h)
    喷嘴(组)名义流量
    (m3/h)
    1~71278128
    1~82559256
    1~951110512
    1~10102311/12/13/14/15/16/17/18/19/20/21/22/23/24/25/261024
    2-6-7-8-11*125027/28/291250
    *注:11号喷嘴可以替换为编号为12~26中的任意一只喷嘴。
    下载: 导出CSV

    表  4  装置1核查数据

    Table  4.   Intermediate check data of device 1

    喷嘴(组)实际流量 (m3/h)xi (1/m3Urel,ik=2)$\dfrac{ {\left| { {x_i} - {x_{\text{r} } } } \right|} }{ { {x_{\text{r} } } } } \times 100{\text{%}}$ 喷嘴(组)实际流量 (m3/h)xi (1/m3Urel,ik=2)$\dfrac{ {\left| { {x_i} - {x_{\text{r} } } } \right|} }{ { {x_{\text{r} } } }} \times 100{\text{%}}$
    1~7124.775251.4570.31%0.024% /////
    8125.765252.7210.31%0.024%181002.215260.5960.31%0.009%
    1~8250.515247.2570.31%0.140%191001.305262.1020.31%0.038%
    9251.945254.6120.31%0.140%201002.045260.2990.31%0.003%
    1~9502.185248.3170.31%0.128%211001.535255.8630.31%0.081%
    10502.985255.0560.31%0.128%221000.505263.0910.31%0.056%
    1~101002.875256.0890.31%0.077%231001.595260.8060.31%0.013%
    111002.095256.4030.31%0.071%241001.735261.1960.31%0.020%
    121002.835259.3790.31%0.014%251001.635262.2440.31%0.040%
    131000.295260.2100.31%0.002%261001.945263.2910.31%0.060%
    141002.015260.4830.31%0.007%271215.895261.3810.31%0.022%
    151001.405260.0470.31%0.001%281217.405265.7180.31%0.061%
    161001.555259.9090.31%0.004%291221.075264.3720.31%0.035%
    171002.295260.0370.31%0.002%2-6-7-8-111220.865258.5800.31%0.075%
    下载: 导出CSV

    表  5  3套装置的核查数据

    Table  5.   Intermediate check data of 3 devices

    测量设备${y_i}$(1/m3相对扩展不确定度Urel,ik=2)$\dfrac{ { {y_i} - \bar y} }{ {\bar y} } \times 100{\text{%}}$$\sqrt {\dfrac{2}{3} } {U_{ {\text{rel} },i} }$核查结论
    装置15267.7130.31%0.12%0.25%核查通过
    装置25265.8480.31%0.08%0.25%核查通过
    装置35251.3090.29%0.20%0.24%核查通过
    下载: 导出CSV
  • [1] 国家市场监督管理总局, 国家标准化管理委员会. 检测和校准实验室能力的通用要求: GB/T 27025—2019[S]. 北京: 中国标准出版社, 2019.
    [2] 乔新愚. 期间核查常见错误解析[J]. 焦作大学学报, 2021, 35(3): 82-84.
    [3] 国家质量监督检验检疫总局. 法定计量检定机构考核规范: JJG1069—2012[S]. 北京: 中国标准出版社, 2012.
    [4] 苏彦勋, 梁国伟, 盛健. 流量计量与测试[M]. 第2版. 北京: 中国计量出版社, 2007: 235.
    [5] 陈炜骄, 孟涛, 李晓鹏. 水流量标准装置期间核查方法研究[J]. 计量技术, 2020(6): 76-79.
    [6] 中国合格评定国家认可委员会. 测量设备期间核查的方法指南: CNAS-GL042: 2019[S]. 北京: 中国标准出版社, 2019.
    [7] 郭爱华. 标准表法气体流量标准装置[J]. 自动化仪表, 2009, 30(8): 50-54. doi: 10.3969/j.issn.1000-0380.2009.08.017
    [8] 李鹏, 刘锴, 张硕, 等. 一种音速喷嘴气体流量标准装置期间核查方法[J]. 中国计量, 2017(1): 105-106.
    [9] 李鹏, 暴冰, 李春辉. 浅析音速喷嘴装置中喷嘴间的相互影响[J]. 计测技术, 2015, 35(5): 52-56. doi: 10.11823/j.issn.1674-5795.2015.05.12
    [10] 全国法制计量管理计量技术委员会. JJF1033-2016《计量标准考核规范》实施指南[M]. 北京: 中国质检出版社, 2017: 118-119.
    [11] 国家质量监督检验检疫总局. 临界流文丘里喷嘴法气体流量标准装置校准规范: JJF 1240—2010[S]. 北京: 中国标准出版社, 2010.
    [12] 国家质量监督检验检疫总局. 涡轮流量计检定规程: JJG1037—2008[S]. 北京: 中国标准出版社, 2008.
    [13] 陈成仁, 王永泉, 施昌彦, 等. 对实验室设备期间核查的理解和应用[J]. 中国计量, 2009(6): 41-43. doi: 10.3969/j.issn.1006-9364.2009.06.027
    [14] 金象春, 肖才锦, 姚永刚, 等. 中子活化分析测定大气颗粒物的质量保证[J]. 计量科学与技术, 2022, 66(2): 9-14.
    [15] 孟成林, 邢静芳. 高等级气体流量标准装置检测能力验证[J]. 工业计量, 2021, 31(2): 36-38.
    [16] 周晓亮, 张立谦, 刘锴, 等. 临界流文丘里喷嘴湿度修正因子的研究[J]. 计量技术, 2019(4): 15-18.
    [17] 黎荣发, 凌光盛, 赵豪, 等. 低压大流量热式气体质量流量计分流测试方法研究[J]. 计量科学与技术, 2022, 66(8): 3-6,12.
    [18] 裴全斌, 青青, 陈正文, 等. 天然气流量计算机计量测试评价方法探讨[J]. 计量科学与技术, 2021, 65(12): 35-39. doi: 10.12338/j.issn.2096-9015.2020.0246
    [19] 杨文武, 贺敬良, 车跃跃. 音速喷嘴式气体流量标准装置试验分析[J]. 计量与测试技术, 2022, 49(5): 36-40. doi: 10.15988/j.cnki.1004-6941.2022.5.012
  • 加载中
图(1) / 表(5)
计量
  • 文章访问数:  469
  • HTML全文浏览量:  167
  • PDF下载量:  56
  • 被引次数: 0
出版历程
  • 收稿日期:  2022-04-01
  • 录用日期:  2022-04-18
  • 网络出版日期:  2023-02-09
  • 刊出日期:  2023-01-18