麻醉深度监测设备检测方法研究

解涛瑞 马雪然 陈达 梁俊辉 定翔

解涛瑞,马雪然,陈达,等. 麻醉深度监测设备检测方法研究[J]. 计量科学与技术,2024, 68(5): 32-36 doi: 10.12338/j.issn.2096-9015.2024.0016
引用本文: 解涛瑞,马雪然,陈达,等. 麻醉深度监测设备检测方法研究[J]. 计量科学与技术,2024, 68(5): 32-36 doi: 10.12338/j.issn.2096-9015.2024.0016
XIE Taorui, MA Xueran, CHEN Da, LIANG Junhui, DING Xiang. Research on Calibration Methods for Anesthetic Depth Monitoring Devices[J]. Metrology Science and Technology, 2024, 68(5): 32-36. doi: 10.12338/j.issn.2096-9015.2024.0016
Citation: XIE Taorui, MA Xueran, CHEN Da, LIANG Junhui, DING Xiang. Research on Calibration Methods for Anesthetic Depth Monitoring Devices[J]. Metrology Science and Technology, 2024, 68(5): 32-36. doi: 10.12338/j.issn.2096-9015.2024.0016

麻醉深度监测设备检测方法研究

doi: 10.12338/j.issn.2096-9015.2024.0016
基金项目: 国家重点研发计划项目(2022YFF0606102)。
详细信息
    作者简介:

    解涛瑞(1999-),中国计量科学研究院在读研究生,研究方向:医学计量,邮箱:Tory.R.Xie@outlook.com

    通讯作者:

    定翔(1984-),中国计量科学研究院副研究员,研究方向:医学计量,邮箱:dingxiang@nim.ac.cn

  • 中图分类号: TB99

Research on Calibration Methods for Anesthetic Depth Monitoring Devices

  • 摘要: 针对麻醉深度监测设备计量检测和性能评价方法缺失的问题,提出采用临床采集的真实脑电信号评价麻醉深度指数测量结果的方法。利用自制麻醉深度监测设备检测系统,将青年和老年患者的麻醉状态脑电信号输入BIS指数、CSI指数、Ai指数等三种麻醉深度监测设备,对测量结果的一致性和重复性进行了计算和评估,分析了指数与脑电信号频谱的相关性。三种指数在反映麻醉状态变化的整体趋势上有较好的一致性,但在麻醉过渡状态和老年患者样本的情况下表现出显著差异。结果表明该方法可用于三种麻醉深度监测设备的检测,对开展麻醉深度监测设备计量质控具有一定的指导意义。
  • 图  1  不同麻醉状态下脑电波图

    Figure  1.  EEG waveforms under different anesthetic states

    图  2  麻醉术中患者脑电频谱的变化

    Figure  2.  Changes in patients’ EEG spectrum during anesthesia

    图  3  浅麻状态下年轻患者与老年患者的脑电信号

    Figure  3.  EEG signals of young and elderly patients under light anesthesia

    图  4  麻醉深度监测设备与信号发生装置连接方式

    Figure  4.  Connection between the anesthetic depth monitoring device and the signal generating device

    图  5  青年患者麻醉深度指数变化

    Figure  5.  Changes in the anesthetic depth index of a young patient

    图  6  老年患者麻醉深度指数变化

    Figure  6.  Changes in the anesthetic depth index of an elderly patient

    图  7  三种麻醉深度指数连续测量的示值波动线图

    Figure  7.  Error line graph for continuous measurements of three anesthetic depth indices

    图  8  年轻患者(左)与老年患者(右)麻醉过程中脑电信号频谱

    Figure  8.  EEG signal spectrum during anesthesia in young patients (left) and elderly patients (right)

  • [1] 定翔, 马雪然, 张杨, 等. 麻醉气体的红外吸收光谱等效模拟方法[J]. 计量学报, 2023, 44(10): 1626-1630. doi: 10.3969/j.issn.1000-1158.2023.10.20
    [2] 张亚妮, 阎文军. 麻醉意识指数在临床麻醉深度监测中的研究进展[J]. 中国临床新医学, 2021, 14(12): 1242-1246. doi: 10.3969/j.issn.1674-3806.2021.12.19
    [3] 邝鹰, 肖暨艳. 麻醉深度监测方法的进展[J]. 现代仪器与医疗, 2016, 22(2): 8-10. doi: 10.11876/mimt201602003
    [4] Brown E N, Lydic R, Schiff N D. General Anesthesia, Sleep, and Coma[J]. New England Journal of Medicine, 2010, 363(27): 2638-2650. doi: 10.1056/NEJMra0808281
    [5] Rampil I J. A Primer for EEG Signal Processing in Anesthesia[J]. Anesthesiology, 1998, 89(4): 980-1002. doi: 10.1097/00000542-199810000-00023
    [6] Bruhn J, Myles P S, Sneyd R, et al. Depth of anaesthesia monitoring: what’s available, what’s validated and what’s next?[J]. British Journal of Anaesthesia, 2006, 97(1): 85-94. doi: 10.1093/bja/ael120
    [7] Bowdle T A. Depth of Anesthesia Monitoring[J]. Anesthesiology Clinics of North America, 2006, 24(4): 793-822. doi: 10.1016/j.atc.2006.08.006
    [8] 曹桂林, 张建欣, 张传阳, 等. 监测麻醉深度对老年患者术中麻醉药用量及术后恢复的影响[J]. 国际麻醉学与复苏杂志, 2017, 38(3): 221-272. doi: 10.3760/cma.j.issn.1673-4378.2017.03.007
    [9] 方建新. 一种用于麻醉深度监测仪的CSI指数提取方法: ZL201610293066.3[P]. 2016-07-20.
    [10] Anderson R E, Jakobsson J G. Cerebral state monitor, a new small handheld EEG monitor for determining depth of anaesthesia: a clinical comparison with the bispectral index during day-surgery[J]. European Journal of Anaesthesiology | EJA, 2006, 23(3): 208.
    [11] Lobo F A, Saraiva A P, Nardiello I, et al. Electroencephalogram Monitoring in Anesthesia Practice[J]. Current Anesthesiology Reports, 2021, 11(3): 169-180. doi: 10.1007/s40140-021-00461-6
    [12] Colombo M A, Napolitani M, Boly M, et al. The spectral exponent of the resting EEG indexes the presence of consciousness during unresponsiveness induced by propofol, xenon, and ketamine[J]. NeuroImage, 2019, 189: 631-644. doi: 10.1016/j.neuroimage.2019.01.024
    [13] Purdon P L, Sampson A, Pavone K J, et al. Clinical Electroencephalography for Anesthesiologists Part I: Background and Basic Signatures[J]. Anesthesiology, 2015, 123(4): 937-960. doi: 10.1097/ALN.0000000000000841
    [14] Sebel P S, Bowles S M, Saini V, et al. Eeg bispectrum predicts movement during thiopental/isoflurane anesthesia[J]. Journal of Clinical Monitoring, 1995, 11(2): 83-91. doi: 10.1007/BF01617729
    [15] Johansen J W. Update on Bispectral Index monitoring[J]. Best Practice & Research Clinical Anaesthesiology, 2006, 20(1): 81-99.
    [16] 珠淮, 刘军. 一种脑电信号监测处理方法: ZL201710076293.5[P]. 2017-07-14.
    [17] 刘军, 周雅琪, 陈绍宾, 等. 基于样本熵与决策树的麻醉意识深度评价指数的研究[J]. 生物医学工程学杂志, 2015, 32(2): 434-439. doi: 10.7507/1001-5515.20150078
    [18] Purdon P L, Pavone K J, Akeju O, et al. The Ageing Brain: Age-dependent changes in the electroencephalogram during propofol and sevoflurane general anaesthesia[J]. BJA: British Journal of Anaesthesia, 2015, 115(Suppl 1): i46-i57.
    [19] 周璞真, 左云霞. 脑电双频指数监测对胃肠外科全麻手术患者麻醉恢复时间影响的回顾性队列研究[J]. 中国循证医学杂志, 2018, 18(6): 565-569. doi: 10.7507/1672-2531.201803105
    [20] Medtronic plc. BISTM MONITORING VALUE RANGE[M/OL].https://www.medtronic.com/content/dam/covidien/library/us/en/product/brain-monitoring/bis-monitoring-system-monitoring-value-range-information-sheet.pdf.
    [21] Fu Y, Xu T, Xie K, et al. Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial[J]. BioMed Research International, 2019, 2019: 1-7. doi: 10.1155/2019/1014825.
    [22] Lobo F A, Vacas S, Rossetti A O, et al. Does electroencephalographic burst suppression still play a role in the perioperative setting?[J]. Best Practice & Research Clinical Anaesthesiology, 2021, 35(2): 159-169.
    [23] Hajat Z, Ahmad N, Andrzejowski J. The role and limitations of EEG-based depth of anaesthesia monitoring in theatres and intensive care[J]. Anaesthesia, 2017, 72(S1): 38-47. doi: 10.1111/anae.13739
    [24] Musialowicz T, Mervaala E, Kälviäinen R, et al. Can BIS monitoring be used to assess the depth of propofol anesthesia in the treatment of refractory status epilepticus?[J]. Epilepsia, 2010, 51(8): 1580-1586. doi: 10.1111/j.1528-1167.2009.02514.x
    [25] Jameson L C, Sloan T B. Using EEG to monitor anesthesia drug effects during surgery[J]. Journal of Clinical Monitoring and Computing, 2006, 20(6): 445-472. doi: 10.1007/s10877-006-9044-x
  • 加载中
图(8)
计量
  • 文章访问数:  155
  • HTML全文浏览量:  35
  • PDF下载量:  14
  • 被引次数: 0
出版历程
  • 收稿日期:  2024-01-16
  • 录用日期:  2024-02-22
  • 修回日期:  2024-05-15
  • 网络出版日期:  2024-05-30