XIAO Yang, FAN Fuyou, LIU Haoran, CHU Taiwei, WANG Qian, LI Mengfei, ZHANG Jian, MOU Tiantian, YUN Mingkai, ZHANG Xiaoli, JIN Sunjun, LIANG Juncheng. Preparation Process and Uniformity Analysis of a PET/CT Medical 68Ge-68Ga Phantom[J]. Metrology Science and Technology, 2021, 65(10): 35-40. doi: 10.12338/j.issn.2096-9015.2020.9044
Citation: XIAO Yang, FAN Fuyou, LIU Haoran, CHU Taiwei, WANG Qian, LI Mengfei, ZHANG Jian, MOU Tiantian, YUN Mingkai, ZHANG Xiaoli, JIN Sunjun, LIANG Juncheng. Preparation Process and Uniformity Analysis of a PET/CT Medical 68Ge-68Ga Phantom[J]. Metrology Science and Technology, 2021, 65(10): 35-40. doi: 10.12338/j.issn.2096-9015.2020.9044

Preparation Process and Uniformity Analysis of a PET/CT Medical 68Ge-68Ga Phantom

doi: 10.12338/j.issn.2096-9015.2020.9044
  • Available Online: 2021-05-08
  • Publish Date: 2021-10-18
  • 68Ge-68Ga phantoms are mainly used in PET/CT image uniformity adjustment and SUV value correction. In order to produce a 68Ge-68Ga phantom, an epoxy resin modified by polysulfide rubber was used as the substrate and the carrier solution of 68Ge-68Ga was used to simulate 68Ge-68Ga radioactive solution. The homogeneous mixture of carrier solution, epoxy resin, and curing agent was filled in a shell of high density polyethylene (HDPE). After the mixture was solidified completely, a non-radioactive phantom with hard texture, color transparent and non-bubble was prepared. Inductively coupled plasma-optical emission spectrometer (ICP-OES) was used to detect the concentration of Ge and Ga elements in the phantom. The results of uniformity analysis showed that there was no significant difference in the concentration distribution of Ge and Ga in the phantom, and the relative standard deviations were 2.20% and 3.65%, respectively, which were equivalent to similar imported products. It was indicated that the process described in this paper can be used for the preparation of 68Ge-68Ga phantom.
  • loading
  • [1]
    D'Angelo E C, Paolisso P, Vitale G, et al. Diagnostic Accuracy of Cardiac Computed Tomography and 18F-Fluorodeoxyglucose With Positron Emission Tomography in Cardiac Masses[J]. JACC: Cardiovascular imaging, 2020, 13(11): 2400-2411. doi: 10.1016/j.jcmg.2020.03.021
    [2]
    Blodgett T M, Meltzer C C, Townsend D W. PET/CT: form and function[J]. Radiology, 2007, 242(2): 360-385. doi: 10.1148/radiol.2422051113
    [3]
    Maurer A H. Combined imaging modalities: PET/CT and SPECT/CT[J]. Health physics, 2008, 95(5): 571-576. doi: 10.1097/01.HP.0000334064.46217.20
    [4]
    Kim Y, Cho K G, Jang S J. Comparison of dual-time point 18F-FDG PET/CT tumor-to-background ratio, intraoperative 5-aminolevulinic acid fluorescence scale, and Ki-67 index in high-grade glioma[J]. Medicine, 2019, 98(8): 14397-14402. doi: 10.1097/MD.0000000000014397
    [5]
    Zhang W S, Ning N, Li X J, et al. Impaired brain glucose metabolism in cirrhosis without overt hepatic encephalopathy: a retrospective 18F-FDG PET/CT study[J]. NeuroReport, 2019, 30(11): 776-782. doi: 10.1097/WNR.0000000000001284
    [6]
    Salmon E, Bernard I C, Hustinx R. Pitfalls and Limitations of PET/CT in Brain Imaging[J]. Seminars in Nuclear Medicine, 2015, 45(6): 541-551. doi: 10.1053/j.semnuclmed.2015.03.008
    [7]
    Chuang H H, Emery D J, Campbell R M, et al. FDG PET/CT in Diagnosing COVID-19 Infection in a Cancer Patient with Exposure History but Minimal Symptoms[J]. Clinical Nuclear Medicine, 2020, 45(8): 656-658. doi: 10.1097/RLU.0000000000003137
    [8]
    钱根年, 陈自谦, 钟群, 等. PET/CT成像系统的质量控制[J]. 中国医疗设备, 2019, 34(2): 1-4. doi: 10.3969/j.issn.1674-1633.2019.02.001
    [9]
    耿建华, 陈英茂, 田嘉禾, 等. 校正PET图像上病灶SUV值的研究[J]. 核电子学和探测技术, 2009, 29(4): 925-929.
    [10]
    万良荣, 黄钢, 孙晓光, 等. PET/CT的质量控制和日常维护[J]. 上海第二医科大学学报, 2005, 25(12): 1303-1305.
    [11]
    Zimmerman B E, Cessna J T. Development of a traceable calibration methodology for solid 68Ge/68Ga sources used as a calibration surrogate for 18F in radionuclide activity calibrators[J]. Journal of Nuclear Medicine Official Publication Society of Nuclear Medicine, 2010, 51(3): 448-453.
    [12]
    Hasford F, Wyk B V, Mabhengu T, et al. Effect of Radionuclide Activity Concentration on PET-CT Image Uniformity[J]. World Journal of Nuclear Medicine, 2016, 15(2): 91-95. doi: 10.4103/1450-1147.167578
    [13]
    Zimmerman B E, Bergeron D E, Cessna J T. Impact of Recent Change in the National Institute of Standards and Technology Standard for 18F on the Relative Response of 68Ge-Based Mock Syringe Dose Calibrator Standards[J]. Journal of Nuclear Medicine Official Publication Society of Nuclear Medicine, 2015, 56(9): 1453-1457.
    [14]
    傅红宇, 罗文博, 李光, 等. 68Ge校正源的制备工艺[J]. 核化学与放射化学, 2013, 35(6): 375-378.
    [15]
    Zimmerman B E, Bergeron D E, Fitzgerald R, et al. Long-term stability of carrier-added Ge-68 standardized solutions[J]. Applied Radiation & Isotopes, 2016(109): 214-216.
    [16]
    王少龙, 柳亚斌, 王瑞山, 等. 二氧化锗离心过滤工艺研究[J]. 稀有金属材料与工程, 2015, 44(8): 1994-1998.
    [17]
    欧阳游, 梁永顺, 唐晓川, 等. 应用方差分析法检验γ测井模型的均匀性[J]. 宇航计测技术, 2018, 38(5): 91-95.