Volume 68 Issue 5
May  2024
Turn off MathJax
Article Contents
HU Hongbo. Improved Nonlinearity Correction Method for Homodyne Orthogonal Signals in Vibration Calibration[J]. Metrology Science and Technology, 2024, 68(5): 3-10, 31. doi: 10.12338/j.issn.2096-9015.2024.0064
Citation: HU Hongbo. Improved Nonlinearity Correction Method for Homodyne Orthogonal Signals in Vibration Calibration[J]. Metrology Science and Technology, 2024, 68(5): 3-10, 31. doi: 10.12338/j.issn.2096-9015.2024.0064

Improved Nonlinearity Correction Method for Homodyne Orthogonal Signals in Vibration Calibration

doi: 10.12338/j.issn.2096-9015.2024.0064
  • Received Date: 2024-03-03
  • Accepted Date: 2024-03-28
  • Rev Recd Date: 2024-05-22
  • Available Online: 2024-05-29
  • Publish Date: 2024-05-18
  • Addressing the issue of nonlinearity correction for homodyne interference signals in vibration calibration, this paper describes an improved ellipse fitting algorithm. Compared to the traditional Heydeman ellipse correction method based on least squares, the proposed method makes two improvements. First, it adds constraints on the ellipse parameters in the fitted conic equation. Second, it transforms the conventional parameter calculation based on the minimum distance criterion into solving the eigenvalues of the data matrix. The proposed method ensures that even when the Lissajous figure formed by the homodyne interference signal is only a partial ellipse, a complete ellipse can still be fitted. This extends the working frequency range of the homodyne interferometer in vibration calibration while improving the measurement accuracy.
  • loading
  • [1]
    ISO. Methods for the calibration of vibration and shock transducers—Part 11: Primary vibration calibration by laser interferometry : ISO 1999 International Standard 16063-11 [S]. Geneva: International Organization for Standardization, 1999.
    [2]
    Marek Dobosz, Takashi Usuda, Tomizo Kurosawa. Methods for the calibration of vibration pick-ups by laser interferometry: I. Theoretical analysis[J]. Meas. Sci. Technol, 1998, 9: 232-239. doi: 10.1088/0957-0233/9/2/010
    [3]
    TaeBong Eom, HongYun Kim, Kyuwon Jeong. The dynamic compensation of nonlinearity in a homodyne laser interferometer[J]. Meas. Sci. Technol, 2001, 12: 1734-1738. doi: 10.1088/0957-0233/12/10/318
    [4]
    Peter L. M. Heydeman. Determination and correction of quadrature fringe measurement errors in interferometers[J]. Appl. Opt, 1981, 20(19): 3382-3384. doi: 10.1364/AO.20.003382
    [5]
    胡红波, 于梅. 零差激光干涉仪正交相位误差的分析[J]. 光学工程, 2012, 39(12): 55-62.
    [6]
    胡红波, 杨丽峰, 于梅. 零差干涉仪用于振动校准中关键技术的研究[J]. 计量学报, 2018, 39(3): 368-372. doi: 10.3969/j.issn.1000-1158.2018.03.16
    [7]
    卢明臻, 高思田, 施玉书, 等. 单频激光干涉仪非线性误差修正方法[J]. 计量学报, 2010, 31(4): 289-293. doi: 10.3969/j.issn.1000-1158.2010.04.01
    [8]
    金涛, 唐一揆, 乐燕芬, 等. 一种单频激光干涉仪非线性误差修正方法研究[J]. 计量学报, 2020, 41(6): 676-681. doi: 10.3969/j.issn.1000-1158.2020.06.08
    [9]
    Keem T, Gonda S, Misumi I, et al. Removing nonlinearity of a homodyne interferometer by adjusting the gains of its quadrature detector systems[J]. Appl. Opt(S0003-6935), 2004, 43(12): 2443-2448.
    [10]
    Rainer Koning, Gejza Wimmer, Viktor Witkovsky. The statistical uncertainty of the Heydemann correction: a practical limit of optical quadrature homodyne interferometry[J]. Meas. Sci. Technol, 2015, 26: 084004. doi: 10.1088/0957-0233/26/8/084004
    [11]
    Hans-Jürgen von Martens. Metrology of Vibration Measurements by Laser Techniques[C]. Eighth International Conference on Vibration Measurements by Laser Techniques: Advances and Applications, 2008.
    [12]
    Alfred Link, Hans-Jürgen von Marterns, Wolfgang Wobinski, et al. Recent advances in vibration and shock measurements and calibrations using laser interferometry[C]. Sixth International Conference on Vibration Measurements by Laser Techniques: advances and applications , 2004.
    [13]
    Sun Q, WabinskiW, Bruns T. Investigation of primary vibration calibration at high frequencies using the homodyne quadrature sine-approximation method: problems and solutions[J]. Meas. Sci. Technol, 2006, 17: 2197-2205. doi: 10.1088/0957-0233/17/8/020
    [14]
    D S Barwick. Very fast best-fit circular and elliptical boundaries by chord data[C]. IEEE Trans. Pattern Anal. Mach. Intell. , 2009.
    [15]
    P Sturm, P Gargallo. Conic fitting using the geometric distance[C]. Asian Conference on Computer Vision, 2007.
    [16]
    张令春, 姜海明, 张俊喜, 等. 局部宜昌因子优化的椭圆拟合算法及其在光纤振动传感器相位解调中的应用[J]. 物理学报, 2022, 71(19): 194206.
    [17]
    Fitzgibbon A, Pilu M, Fisher R B. Direct least square fitting of ellipses[J]. IEEE Trans. Pattern Anal. Mach. Intell, 1999, 21(5): 476-480. doi: 10.1109/34.765658
    [18]
    Harker M, O’Leary P, Zsombor-Murray P. Direct type-specific conic fitting and eigenvalue bias correction[J]. Image Vis. Comput, 2008, 26: 372-381. doi: 10.1016/j.imavis.2006.12.006
    [19]
    胡红波, 于梅. 绝对法冲击校准激光干涉信号处理技术[J]. 计量技术, 2013, 1: 46-50.
    [20]
    夏菁. 椭圆拟合方法的比较研究[D]. 广州: 暨南大学, 2007.
    [21]
    Christiaan S Celdman. Performing accelerometer calibration using homodyne laser interferometry with displacements smaller than λ/4[J]. ACTA IMEKO, 2017, 6(4): 75-79. doi: 10.21014/acta_imeko.v6i4.487
    [22]
    Bev Payne. Calibration of accelerometers at extended frequencies utilizing piezoelectric shakers[C]. Atlanta : 84th Shock and Vibration Symposium, 2013.
    [23]
    于梅, 胡红波, 左爱斌, 等. 新一代冲击加速度国家计量基准装置的研究与建立[J]. 振动与冲击, 2013, 32(10): 124-129. doi: 10.3969/j.issn.1000-3835.2013.10.023
    [24]
    鲁敏, 胡红波. 基于正弦激励的压电加速度计模型参数辨识[J]. 计量学报, 2020, 41(1): 55-59. doi: 10.3969/j.issn.1000-1158.2020.01.11
    [25]
    于梅. 0.1 Hz~50 kHz直线振动幅值和相位国家计量基准系统的研究[J]. 振动与冲击, 2007, 26(7): 54-58. doi: 10.3969/j.issn.1000-3835.2007.07.014
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Figures(10)

    Article Metrics

    Article views (109) PDF downloads(12) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return