60 ~ 250 kV治疗水平X射线空气比释动能量值国内比对

    Comparison of Air Kerma Values of X-Ray at 60 ~ 250 kV Therapeutic Level in China

    • 摘要: 治疗水平电离室剂量计是医院放射治疗用的主要计量器具,也是国家强制检定计量器具,其剂量测量的准确是开展精准放疗的关键。为保证量值准确与统一,2019至2020年中国计量科学研究院作为主导单位组织了全国60 ~ 250 kV X射线空气比释动能的量值比对。共有11个实验室参加此次比对,所有参比实验室均建立60 kV、100 kV、135 kV、180 kV和250 kV的参考辐射质。本次比对采用“花瓣式”比对方式,在辐射场中完成两只传递标准电离室的校准测量,得到不同辐射质下的校准因子及测量不确定度。比对结果表明各参比实验室归一化偏差均小于1,比对结果满意。

       

      Abstract: Radiotherapy level ionization chamber dosimeter is the main measuring instrument for radiotherapy in hospitals, and it is also the national compulsory verification measuring instrument. The accuracy of its dose measurement is the key carrying out precision radiotherapy. To ensure the accuracy and unity of the quantity value, from 2019 to 2020, National Institute of metrology, China, as the leading laboratory, organized the national comparison of the 60 ~ 250 kV X-ray air kerma. A total of 11 laboratories participated in the comparison, and all reference laboratories established 60 kV, 100 kV, 135 kV, 180 kV, and 250 kV reference radiation. The calibration factor and measurement uncertainty under different radiation were obtained by calibrating two transfer standard ionization chambers in the radiation field using the petal comparison method. The comparison results show that the normalized deviation of each reference laboratory is less than 1, and the comparison results are satisfactory.

       

    /