手持式激光诱导击穿光谱仪校准方法研究及不确定度评定

    Research on Calibration Method and Uncertainty Evaluation of Handheld Laser-Induced Breakdown Spectrometer

    • 摘要: 介绍了手持式激光诱导击穿光谱仪的应用及发展,由于目前该类设备尚缺乏相应的计量技术规范,因此提出了手持式激光诱导击穿光谱仪的计量校准方法。主要的计量校准项目包括检出限、示值误差、重复性,以实验数据为支撑,给出了相应的计量性能技术指标;并对示值误差不确定度进行了详细分析,给出了相应的评定实例;对手持式激光诱导击穿光谱仪的计量溯源给出了有效的评价方法。

       

      Abstract: The development and application of handheld laser-induced breakdown spectrometers are introduced in this paper. Due to the lack of corresponding metrological technical specifications for this type of equipment, a metrological calibration method for handheld laser-induced breakdown spectrometers is proposed. The main metrological characteristics include detection limit, indication error, and repeatability. Based on the actual test data, the corresponding technical indicators are given. The uncertainty of indication error is analyzed in detail and the corresponding evaluation example is given. An effective evaluation method for the metrological traceability of handheld laser-induced breakdown spectrometer is given.

       

    /