Experiment and Data Analysis of Digital Adjustable Pipette Based on Different Working Modes
-
摘要: 介绍了数字可调移液器的工作原理和3种工作模式,阐述了数字可调移液器的检定方法和检定步骤,给出了数据分析方法,分析了影响测量结果准确性的影响因素。通过普通模式、粘液模式和分液模式3种工作模式下的容量误差和重复性等指标的对比实验,验证了3种工作模式的优劣。Abstract: This paper introduces the working principle and three working modes of the digital adjustable pipette, expounds the verification method and steps of the digital adjustable pipette, gives the data analysis method, and analyzes the influencing factors of the accuracy of the measurement results. Through the comparative experiment of capacity error and repeatability of three working modes, including ordinary mode, mucus mode and liquid separation mode, the advantages and disadvantages of the three working modes are verified.
-
Key words:
- digital adjustable pipette /
- verification method /
- working mode /
- data analysis
-
表 1 移液器容量允许误差和测量重复性
Table 1. Capacity allowable error and measurement repeatability of pipette
标称容量/ μL 检定点/ μL 容量允许误差 /% 测量重复性 /% 1 0.1 ±20.0 10.0 0.5 ±20.0 10.0 1 ±12.0 6.0 2 0.2 ±20.0 10.0 1 ±12.0 6.0 2 ±12.0 6.0 5 0.5 ±20.0 10.0 1 ±12.0 6.0 5 ±8.0 4.0 10 1 ±12.0 6.0 5 ±8.0 4.0 10 ±8.0 4.0 20 2 ±12.0 6.0 10 ±8.0 4.0 20 ±4.0 2.0 25 2 ±12.0 6.0 10 ±8.0 4.0 25 ±4.0 2.0 40 5 ±8.0 4.0 20 ±4.0 2.0 40 ±3.0 1.5 50 5 ±8.0 4.0 25 ±4.0 2.0 50 ±3.0 1.5 100 10 ±8.0 4.0 50 ±3.0 1.5 100 ±2.0 1.0 表 2 检测指标允差表
Table 2. Allowable error table for detection indicators
检定容量/ μL 容量允许误差/% 移液器重复性/% 10 ±8.0 4.0 50 ±3.0 1.5 100 ±2.0 1.0 表 3 普通模式下的误差检测表
Table 3. Error detection table in common mode
检定容量/ μL 实际容量/ μL 容量误差/% 重复性/% 10 9.9 1.0 0.7 50 50.2 −0.4 0.3 100 100.5 −0.5 0.2 表 4 粘液模式下的误差检测表
Table 4. Error detection table in mucus mode
检定容量/ μL 实际容量/ μL 容量误差/% 重复性/% 10 11.7 −14.5 1.7 50 52.6 −4.9 0.2 100 103.0 −2.9 0.5 表 5 分液模式下的误差检测表
Table 5. Error detection table in liquid separation mode
检定容量/ μL 实际容量/ μL 容量误差/% 重复性/% 10 10.7 −6.5 5.5 50 51.6 −3.1 0.7 100 101.3 −1.3 0.5 -
[1] 刘翔, 王湘波, 唐何, 等. 采用质量控制图法对可调移液器进行期间核查的方法研究[J]. 中国药事, 2015, 29(3): 281-286. [2] 佚名. 微量可调移液器[J]. 教学仪器与实验, 2004, 20(8): 43. [3] 汪朝红, 胡良勇, 傅忆宾, 等. 数字可调移液器校准方法研究[J]. 中国测试, 2011, 37(3): 23-25. [4] 李仲贤. 探讨水分蒸发现象对移液器检定数据的影响[J]. 计量与测试技术, 2019, 46(2): 41-43. [5] Justin Beroz, A John Hart. Universal handheld micropipette[J]. Review of Scientific Instruments, 2016, 87: 1-8. [6] 徐晓静, 赵李祥. 浅谈移液器的维护与保养[J]. 实验室科学, 2011, 14(1): 156-157. doi: 10.3969/j.issn.1672-4305.2011.01.051 [7] 吴银笑, 莫怡玉, 吴卓智. 衡量发自校验常用移液器的程序与效果[J]. 仪器仪表与分析监测, 2011(3): 1-5. doi: 10.3969/j.issn.1002-3720.2011.03.001 [8] 马骁勇, 杜淑娟. 移液器检定经验谈[J]. 中国计量, 2006(2): 68-69. doi: 10.3969/j.issn.1006-9364.2006.02.042 [9] 李拥军. 可调移液器的使用与保养[J]. 工业计量, 2007, 17(6): 59. doi: 10.3969/j.issn.1002-1183.2007.06.024 [10] 张珑, 纪建英. 常用玻璃量器测量不确定度评定[J]. 计量技术, 2008(10): 68-69. [11] 李玉梅. 测量设备能力的分析和计算[J]. 计量技术, 1986(4): 8-11. [12] 宋冀红, 侯斌. 定量、可调移液器测量结果的不确定度分析[J]. 计量与测试技术, 2008, 35(7): 51-55. doi: 10.3969/j.issn.1004-6941.2008.07.028