Research on the Effect of Humidity on the Metering Characteristics of Soap Film Flowmeter
-
摘要: 皂膜流量计是开展环境监测工作的重要计量器具。为满足环境监测用流量装置低压试验微小气体流量的有效溯源,减小气体湿度对皂膜流量计流量精确测量的影响,提升微小气体流量量值传递的可靠性,分析了皂膜流量计的工作特性,给出了皂膜流量计的微小气体流量湿度修正计算方法。基于中国计量科学研究院的微小气体流量标准装置进行了实验验证,结果表明该湿度修正计算方法可以有效降低湿度对皂膜流量计计量特性的影响。Abstract: Soap film flowmeter is an important measuring instrument for environmental monitoring. In order to meet the effective traceability of micro gas flow in low-pressure test of flow meter for environmental monitoring, reduce the impact of gas humidity on the accurate measurement of soap film flowmeter, and improve the reliability of micro gas flow value transmission. In this paper, the operating characteristics of soap film flowmeter are analyzed, and the calculation method of humidity correction for micro gas flow is given. The results show that the humidity correction calculation method can effectively reduce the influence of humidity on the metering characteristics of soap film flowmeter.
-
表 1 基于湿度修正后的皂膜流量计的干空气体积
Table 1. The volume of dry air measured by soap film flowmeter based on humidity correction
示值/ml/min 标准容积/ml 时间/s 计算流量/ml/min 修正系数 标准流量/ml/min 示值误差/% 197 200 57.18 210 0.9584 201 −1.99 197 200 57.19 210 0.9584 201 −1.99 309 600 109.66 328 0.9584 314 −1.59 309 600 109.85 328 0.9584 314 −1.59 503 1000 117.1 512 0.9584 491 2.44 503 1000 117.11 512 0.9584 491 2.44 598 400 38.25 627 0.9584 601 −0.5 598 400 38.28 627 0.9584 601 −0.5 698 1000 82.5 727 0.9584 697 0.14 700 800 65.88 728 0.9584 698 0.29 802 1000 72.59 826.5 0.9584 792 1.26 802 1000 72.55 827 0.9584 792 1.26 900 1000 64.86 925 0.9584 886 0.45 900 1000 64.82 926 0.9584 887 0.34 1007 1000 58.04 1034 0.9584 991 1.61 1000 1000 58.69 1022 0.9584 980 2.0 表 2 气路切换到干式活塞流量计Defender 530(50-5000) mL/min的测量数据
Table 2. Measurement data of gas circuit switching to dry piston flowmeter defender 530
示值/mL/min 标准流量/mL/min 示值误差/% 197 198 −0.5 309 311 −0.64 503 492 2.23 598 598 0 697 697 0 802 795 0.88 900 892 0.9 1007 1007 0 -
[1] 徐亚. 气体小流量计量方法研究与标准装置开发[D]. 长沙: 中南大学, 2012. [2] Liu J, Olsson G, Mattiasson B. A volumetric meter for monitoring of low gas flow rate from laboratory-scale biogas reactors[J]. Sensors and Actuators B: Chemical, 2004, 97(2-3): 369-372. doi: 10.1016/j.snb.2003.09.014 [3] 高山, 李春辉, 李小亭. 微小pVTt法气体流量标准装置的性能评估及验证[J]. 中国测试, 2019, 45(3): 112-117. [4] 李培晶. 被动式活塞气体小流量标准装置的研制[D]. 北京: 中国计量科学研究院, 2020. [5] 张书令, 许红, 洪扁, 等. 正压标准漏孔校准装置[J]. 中国测试, 2015(S1): 6-9. [6] 单保超. 皂膜流量计的误差分析与改进设计[J]. 科技创新导报, 2015(3): 1-2. doi: 10.3969/j.issn.1674-098X.2015.03.001 [7] 陆凯荪. 皂膜流量计测量气体小流量及其误差分析[J]. 分析仪器, 1980(3): 86-89. [8] 郭美荣, 张辉. 持续润湿与测量区间对自产泡皂膜流量计的影响研究[J]. 自动化仪表, 2019, 40(4): 73-76. [9] 李建鹏. 电子皂膜流量计使用中问题及建议[J]. 工业计量, 2016(6): 91-93. [10] 朱碧玉. 皂膜流量计校准中湿度的修正[J]. 上海计量测试, 2018(2): 43-47. doi: 10.3969/j.issn.1673-2235.2018.02.016 [11] 朱碧玉, 陈超, 张进明. 皂膜流量计校准中湿度的修正[J]. 上海计量测试, 2018, 45(2): 50-51. [12] 廖丽华. 电子皂膜流量计的研究与设计[D]. 长春: 吉林大学, 2012. [13] 陈汉松, 万勇, 邹昌活, 等. 一种校准微小流量电子皂膜流量计的新方法[J]. 计量技术, 2016(12): 51-53. doi: 10.3969/j.issn.1000-0771.2016.12.15