Research on Flow Parameter Calibration Device of Hemodialysis Equipment Tester
-
摘要: 介绍了血液透析机检测仪的工作原理,对血液透析机检测仪中流量参数的校准方法进行了探讨,研究设计了用于血液透析机检测仪流量参数校准的小流量校准装置。以静态质量法液体流量装置为基础,结合标准流量计作为计量标准器,对血液透析机检测仪流量参数进行校准以及对标准流量计进行标定,并对装置的不确定度进行分析,结果满足流量参数的校准要求。Abstract: This paper introduces the working principle of hemodialysis equipment tester, discusses the calibration method of the flow parameters in the hemodialysis equipment tester, studies and designs a small flow calibration device for the flow parameters calibration of the hemodialysis equipment tester, which is based on the liquid flow device of the static mass method, combined with the standard flowmeter as the measurement standard, and realizes the calibration of hemodialysis equipment tester calibration of flow parameters, the results show that calibration of standard flowmeter and analysis of the uncertainty of the device can meet the calibration requirements of flow parameters.
-
Key words:
- hemodialysis equipment tester /
- flow /
- calibration device /
- calibration methods /
- uncertainty
-
表 1 计时器检测数据
Table 1. Time-meter test data
单位:s 最短测量时间tmin=60 次数 被检时间ti 标准时间t0i 时间差Δti 平均时间差Δt 1 60.6274 60.62769 −0.00026 −0.00015 2 60.6686 60.66866 −0.00009 3 60.9189 60.91901 −0.00016 4 60.7080 60.70804 −0.00004 5 60.7920 60.79213 −0.00013 6 60.3763 60.37651 −0.00022 7 60.8357 60.83586 −0.00014 8 60.3720 60.37211 −0.00011 9 60.5743 60.57443 −0.00014 10 60.4149 60.41502 −0.00016 表 2 电子天平检测数据
Table 2. Electronic balance test data
单位:g 空容器负载平均值:1000 g 次数 加载点 加载砝码m 天平显示R 示值误差Δm 1 1 500 1500.2 0.2 2 1000 2000.4 0.4 3 1500 2500.6 0.6 4 2000 3000.6 0.6 5 2500 3500.4 0.4 6 3000 4000.6 0.6 7 3500 4500.7 0.7 8 4000 5000.5 0.5 9 4500 5500.6 0.6 10 5000 6000.8 0.8 2 10 5000 6000.8 0.8 9 4500 5500.5 0.5 8 4000 5000.5 0.5 7 3500 4500.7 0.7 6 3000 4000.6 0.6 5 2500 3500.5 0.5 4 2000 3000.4 0.4 3 1500 2500.6 0.6 2 1000 2000.3 0.3 1 500 1500.3 0.3 … … … … … 10 10 5000 6000.9 0.9 9 4500 5500.8 0.8 8 4000 5000.8 0.8 7 3500 4500.7 0.7 6 3000 4000.6 0.6 5 2500 3500.6 0.6 4 2000 3000.5 0.5 3 1500 2500.5 0.5 2 1000 2000.3 0.3 1 500 1500.4 0.4 -
[1] 王池, 王自和, 张宝珠, 等. 流量测量技术全书[M]. 北京: 化学工业出版社, 2012. [2] 叶镇国. 水力学与桥涵水文[M]. 北京: 人民交通出版社, 2003. [3] 孙劼, 张璞, 万国庆, 等. 血液透析机检测仪校准装置的研制[J]. 医疗卫生装备, 2019, 40(5): 12-16. [4] 葛剑徽, 滕越, 张玉明, 等. 血液透析装置检测仪流量校准方法研究与应用[J]. 医疗卫生装备, 2014, 35(12): 11-14. [5] 于陆军, 赵伟, 舒心, 等. 液体流量标准装置稳定性计算方法的实验研究与分析[J]. 计量技术, 2020, 64(1): 28-31. [6] 陈炜骄, 孟涛, 李晓鹏. 水流量标准装置期间核查方法研究[J]. 计量技术, 2020, 64(6): 76-79. doi: 10.3969/j.issn.1000-0771.2020.06.18 [7] 李海洋, 陈琪, 姚新红, 等. 高准确度静态质量法液体流量标准装置及其不确定度评价[J]. 计量与测试技术, 2020, 47(5): 65-68. [8] 李高峰, 王宏伟, 陈娟, 等. 水流量标准装置的设计与应用[J]. 计量技术, 2017, 61(8): 49-50. [9] 郭均瑶, 洪传文, 许建平. 液体流量标准装置称重容器的设计[J]. 计量技术, 2017, 61(12): 19-22. [10] 信彦峰, 桑培勇, 贾正红, 等. 液体流量标准装置称重换向系统设计探讨[J]. 计量技术, 2016, 60(12): 61-63. [11] 曲志勇, 王宪民, 胡生伟. 血液透析装置流量监控示值误差的测量不确定度评定[J]. 计量与测试技术, 2014, 41(5): 70-72. [12] 刘敏, 杨焕诚. 静态质量法微小水流量标准装置不确定度评定报告[J]. 内蒙古科技与经济, 2014(16): 76. [13] 郑朴皓. 费森尤斯4008型血液透析机流量与电导度故障分析与处理[J]. 医疗装备, 2015(3): 120-121. doi: 10.3969/j.issn.1002-2376.2015.03.069 [14] 全国医学计量技术委员会. 血液透析装置检测仪校准规范: JJF 1541-2015[S]. 北京: 中国质检出版社, 2015. [15] 全国流量容量技术委员会. 液体流量标准装置检定规程: JJG 164-2000[S]. 北京: 中国质检出版社, 2000.