Uncertainty Evaluation of Different Quantitative Calibration Methods for the Determination of Vitamin B12 in Infant Formula
-
摘要: 本文以婴儿配方奶粉中的维生素B12测量为例,开展了标准加入法、标准加入同位素稀释质谱法、单点同位素稀释质谱法和校准曲线同位素稀释质谱法四种定量方法的不确定度评估。基于ISO-GUM指南等不确定度评估方法对四种定量方法进行不确定度量化评估,发现方法重复性和曲线拟合引入的不确定度为主要不确定度来源,且标准加入法评估得到的不确定度明显大于同位素内标法评估得到的不确定度。通过不同定量方法的比较研究,可以更有效地选择定量方式,提高测量精度。Abstract: In this study, take measurement of the vitamin B12 in infant formula as an example, measurement uncertainty was evaluated through four calibration methods, including standard addition method, standard addition-isotope dilution mass spectrometry method, single point isotope dilution mass spectrometry method, and standard calibration curve isotope dilution mass spectrometry method. The uncertainty evaluation of the four quantitative methods was assessed based on the ISO-GUM guidelines and other uncertainty assessment methods. The uncertainties introduced by the method of repeatability and curve fitting were the primary sources of uncertainty. Furthermore, the uncertainties obtained by the standard addition method were significantly more significant than those obtained by the isotope internal standard method. The comparative study of different quantification methods can lead to a more effective selection of quantification methods and improved measurement accuracy.
-
表 1 不同定量方法相关分量相对标准不确定度
Table 1. Relative standard uncertainty values of each components of the quantitative methods
不确定度来源 SA SA-IDMS SP-IDMS SCC-IDMS A类不确定度 重复性uref,rep 3.3% 1.1% 1.6% 1.3% B类不确定度 标准物质纯度uref,p 0.4% 0.4% 0.4% 0.4% 溶液制备与称量uref,w 0.1% 0.1% 0.1% 0.2% 校准曲线urel,cal 6.3% 2.5% / 2.6% 合成不确定度 7.2% 2.8% 1.7% 3.0% 扩展不确定度U(k=2) 14.4% 5.6% 3.4% 6.0% -
[1] 国家质量监督检验检疫总局. 测量不确定度评定与表示: JJF1059.1-2012[S]. 北京: 中国质检出版社, 2013. [2] DJELLOULI O, ROZET E, DAHMANI B, et al. Compromise in uncertainty estimation by modelling and validation approaches for an HPLC-UV method for measurement of biochemical indicators of vitamins A and E[J]. Analyst, 2018, 143(6): 1492-1501. doi: 10.1039/C7AN01804E [3] KIM J M, CHOI S H, SHIN G H, et al. Method validation and measurement uncertainty for the simultaneous determination of synthetic phenolic antioxidants in edible oils commonly consumed in Korea[J]. Food Chem, 2016, 213: 19-25. doi: 10.1016/j.foodchem.2016.06.053 [4] LI Y, GILL B D, GRAINGER M N C, et al. The analysis of vitamin B12 in milk and infant formula: A review[J]. Int Dairy J, 2019, 99: 104543. doi: 10.1016/j.idairyj.2019.104543 [5] LI X Q, YANG Z, ZHANG Q H, et al. Evaluation of matrix effect in isotope dilution mass spectrometry based on quantitative analysis of chloramphenicol residues in milk powder[J]. Anal Chim Acta, 2014, 807: 75-83. doi: 10.1016/j.aca.2013.11.017 [6] RAPOSO F, BARCELÓ D. Challenges and strategies of matrix effects using chromatography-mass spectrometry: An overview from research versus regulatory view points[J]. TrAC Trend Anal Chem, 2021, 134: 116068. doi: 10.1016/j.trac.2020.116068 [7] BIENVENU JF, PROVENCHER G, BELANGER P, et al. Standardized Procedure for the Simultaneous Determination of the Matrix Effect, Recovery, Process Efficiency, and Internal Standard Association[J]. Anal Chem, 2017, 89(14): 7560-7568. doi: 10.1021/acs.analchem.7b01383 [8] MARTINS ML, RIZZETTI TM, KEMMERICH M, et al. Dilution standard addition calibration: A practical calibration strategy for multiresidue organic compounds determination[J]. J Chromatogr A, 2016, 1460: 84-91. doi: 10.1016/j.chroma.2016.07.013 [9] PÉREZ-FERNÁNDEZ V, GENTILI A, MARTINELLI A, et al. Evaluation of oxidized buckypaper as material for the solid phase extraction of cobalamins from milk: Its efficacy as individual and support sorbent of a hydrophilic-lipophilic balance copolymer[J]. J Chromatogr A, 2016, 1428: 255-266. doi: 10.1016/j.chroma.2015.07.109 [10] ZIRONI E, GAZZOTTI T, BARBAROSSA A, et al. Technical note: development and validation of a method using ultra performance liquid chromatography coupled with tandem mass spectrometry for determination of vitamin B12 concentrations in milk and dairy products[J]. J Dairy Sci, 2013, 96(5): 2832-2836. doi: 10.3168/jds.2012-6451 [11] LUO X, CHEN B, DING L, et al. HPLC-ESI-MS analysis of Vitamin B12 in food products and in multivitamins-multimineral tablets[J]. Anal Chim Acta, 2006, 562(2): 185-189. doi: 10.1016/j.aca.2006.01.073 [12] LIM YO, KIM B, AHN S, et al. Improvement of accuracy for the determination of vitamin A in infant formula by isotope dilution-liquid chromatography/tandem mass spectrometry[J]. Food Chem, 2011, 126(3): 1393-1398. doi: 10.1016/j.foodchem.2010.11.118 [13] SHEHATA AB, RIZK MS, FARAG AM, et al. Development of two reference materials for all trans-retinol, retinyl palmitate, alpha- and gamma-tocopherol in milk powder and infant formula[J]. J Food Drug Anal, 2015, 23(1): 82-92. doi: 10.1016/j.jfda.2014.10.004 [14] LEE J, JANG ES, KIM B. Development of isotope dilution-liquid chromatography/mass spectrometry combined with standard addition techniques for the accurate determination of tocopherols in infant formula[J]. Anal Chim Acta, 2013, 787: 132-139. doi: 10.1016/j.aca.2013.05.034 [15] ISO. Guide to the Expression of Uncertainty in Measurement: ISO 1993 (E)[S]. ISO, 1993. [16] 中国合格评定国家认可委员会. 化学分析中不确定度的评估指南: CNAS-GL006[S]. 北京: 中国计量出版社, 2019. [17] 孔祥明, 魏晓琴. 不同定量方法对蔬菜农药残留测量不确定度的影响研究[J]. 宁夏农林科技, 2020, 61(11): 34-37. [18] 田甜, 刘书妤, 张燕, 等. 蜂蜜中氯霉素痕量残留测量质谱基质效应研究[J]. 计量科学与技术, 2021, 65(5): 77-82,100. doi: 10.12338/j.issn.2096-9015.2020.9037 [19] LIM S, YUN C I, KO K Y, et al. Validation and measurement uncertainty of HPLC method for simultaneous determination of 10 dyes in adulterated Phellodendron[J]. Korean J Food Sci Technol, 2021, 53(4): 391-399.