Establishment of Metrological Traceability System for Mass Concentration of Ambient Air Particle
-
摘要: 针对环境空气颗粒物浓度测量中的测不准、难溯源的难题,紧密结合国家标准以及国家计量发展规划对环境空气中气溶胶颗粒物质量浓度计量溯源的需求,对颗粒物采样、颗粒物粒径分离和颗粒物质量浓度测量等关键技术,全方位系统地叙述了关键技术中国家标准物质、计量标准装置、技术国际等效性和国家标准技术攻关的相关情况,立体搭建了环境空气颗粒物溯源体系。保障了环境空气颗粒物监测数据的准确可靠,以及与国际同类监测数据的一致性,建立了我国环境空气颗粒物监测溯源基准。Abstract: In view of the difficulties of uncertainty and traceability in the measurement of ambient air particle concentration, and in close combination with the requirements of national standards and Metrological Development Plans for measurement and traceability of aerosol particle mass concentration, the key technologies such as particle sampling, particle size separation and particle mass concentration measurement are focused on, and the metrological traceability system for ambient air particles has been established through scientific and technological research, which is carried out systematically on national reference materials, measurement standard devices, international equivalence of technologies and national standards. The project research ensures the accuracy and reliability of the monitoring data of ambient air particles, the consistency with international monitoring data and the traceability standard’s establishment in the field in China.
-
表 1 环境空气颗粒物监测仪用标准物质
Table 1. Reference materials for ambient air particle monitor
名称 编号 单位 空气动力学
当量直径不确
定度PM2.5监测仪检
测用标准物质GBW13642 nm 1491 38 GBW13643 1837 41 GBW13644 2217 45 GBW13645 2505 49 GBW13646 2813 52 GBW13647 3216 57 GBW13648 3748 64 GBW13649 4220 71 PM10监测仪检
测用标准物质GBW13656 μm 3.02 0.05 GBW13657 4.91 0.07 GBW13658 7.29 0.10 GBW13659 9.27 0.13 GBW13660 11.20 0.16 GBW13661 13.36 0.18 GBW13662 15.25 0.22 GBW13663 17.34 0.24 PM10切割器切
割特性监测用
标准物质GBW(E)130575 μm 4.65 0.14 GBW(E)130576 7.37 0.17 GBW(E)130577 9.22 0.17 GBW(E)130578 10.57 0.17 GBW(E)130579 12.52 0.19 GBW(E)130580 14.63 0.20 GBW(E)130581 18.56 0.23 -
[1] EPA. Environmental Protection Agency 40 CFR Part 50: National primary and secondary ambient air quality standards [S]. US: EPA, 2006. [2] EPA. Environmental Protection Agency 40 CFR Parts 53 and 58: Revisions to Ambient Air Monitoring Regulations [S]. US: EPA, 2006. [3] European Committee for Standardization. BS EN 14907-2005: Ambient air quality. Standard gravimetric measurement method for the determination of the PM2.5 mass fraction of suspended particulate matter [S]. UK: BSI, 2005. [4] 张文阁, 刘巍, 许潇, 等. PM2.5监测仪检测用国家一级标准物质的研制[J]. 计量学报, 2019, 40(1): 159-163. [5] 张文阁, 周文刚, 许潇, 等. PM10监测仪检测用国家一级标准物质的研制[J]. 中国计量, 2020(2): 83-85. [6] 张文阁. PM2.5切割器切割特性检测用标准物质的定值方法及测量系统: CN201611104452. X[P]. 2020-02-14. [7] 张文阁, 许潇. PM2.5切割器切割特性检测装置: ZL201720651780.5[P]. 2017-12-12. [8] 陈仲辉, 张文阁, 黄星亮, 等. PM2.5切割器切割特性研究[J]. 中国计量, 2014(8): 82-85. [9] 张文阁, 高思田, 宋小平, 等. 细颗粒物PM2.5浓度测量及计量技术[J]. 中国粉体技术, 2013, 19(6): 69-72. [10] 张文阁, 许潇. PM2.5切割器D50不确定度的测定方法及系统: ZL201710416703.6[P]. 2017-09-22. [11] 王婷, 刘巍, 张明, 等. 切割器切割特性试验装置的功能性验证方法探讨[J]. 计量科学与技术, 2022, 66(1): 41-45. [12] 张文阁. PM2.5质量浓度监测仪校准装置: ZL201621463092.8[P]. 2017-09-12. [13] 张文阁, 高思田, 宋小平, 等. PM2.5颗粒物采样器校准装置设计及技术关键[J]. 中国粉体工业, 2014(1): 17-18. [14] 张文阁, 刘巍, 陈仲辉, 等. PM2.5重量法标准装置研制及溯源性研究[J]. 计量技术, 2015(1): 3-7. [15] 张文阁. JJF1659-2017《PM2.5质量浓度测量仪校准规范》解读[J]. 中国计量, 2019(2): 89-92,114. [16] 张文阁, 许潇, 刘俊杰. 中国、英国PM2.5质量浓度溯源方法双边比对[J]. 中国计量, 2021(6): 70-73. [17] 国家市场监督管理总局. 环境空气 颗粒物质量浓度测定 重量法: GB/T 39193-2020[S]. 北京: 中国标准出版社, 2020. [18] 国家市场监督管理总局. 颗粒 生物气溶胶采样器 技术条件: GB/T 39990-2021[S]. 北京: 中国标准出版社, 2021. [19] 生态环境部. 环境空气颗粒物(PM10和PM2.5)连续自动监测系统技术要求及检测方法: HJ 653-2021[S]. 北京: 中国环境科学出版社, 2021. [20] 中国颗粒学会, 北京粉体技术协会. 光散射原理细颗粒物(PM2.5)质量浓度测量仪性能测试方法: T/CSP T/BSPT 003-2018[S]. 北京: 中国颗粒学会, 2018. [21] 国家市场监督管理总局. 烟尘采样器: JJG 680-2021[S]. 北京: 中国质检出版社, 2021. [22] 国家质量监督检验检疫总局. PM2.5质量浓度测量仪校准规范: JJF 1659-2017 [S]. 北京: 中国质检出版社, 2017. [23] 国家质量监督检验检疫总局. 大气采样器: JJG 956-2013[S]. 北京: 中国质检出版社, 2013. [24] 国家质量监督检验检疫总局. 烟尘采样器型式评价大纲: JJF 1332-2011[S]. 北京: 中国质检出版社, 2011. [25] 国家质量监督检验检疫总局. 粉尘浓度测量仪: JJG 846-2015[S]. 北京: 中国质检出版社, 2015. [26] 国家市场监督管理总局. 粉尘浓度测量仪型式评价大纲: JJF 1716-2018[S]. 北京: 中国质检出版社, 2018.