Correction and Calibration of R Value Curve for Blood Oxygen Saturation Measurement Model Based on Lambert-Beer Law
-
摘要: 针对血氧模拟仪现有的R值曲线无法满足种类繁多的血氧仪计量检定需求,为解决血氧仪与模拟仪不兼容问题,提出一种基于Lambert-Beer定律的血氧饱和度测量模型的修正方法。在现有的检定装置上添加滤光介质,以实现在原有的测量模型上加一个修正值,从而对R值曲线进行修正。实验表明,通过添加滤光介质修正R值曲线,选择不同R值曲线检测血氧仪时均显示出正常的测量值,且示值重复性较小,符合国家计量检定规程要求,有效解决了血氧模拟仪与血氧仪之间的不兼容问题。
-
关键词:
- Lambert-Beer定律 /
- 血氧饱和度 /
- 模拟仪 /
- R值曲线 /
- 检定
Abstract: The measurement of oxygen saturation in peripheral tissues is an important indicator of the oxygen supply to cells and tissues. However, with the emergence of a variety of new oximeters, the existing R value curve of oximetry simulation can no longer meet the current demand for metrological testing. To address this issue, a modified method of oxygen saturation measurement based on the Lambert-Beer law was proposed. This method involves the addition of an optical filter medium to the existing verification device to correct the R value curve and achieve a correction value added to the original measurement model. The experimental results demonstrate that by adding filter media to modify the R value curve, the oximeter can measure values that are close to the standard value under different R value curves. The indication repeatability is also small, meeting the requirements of national metrological verification regulation. The proposed method effectively solves the incompatibility problem between the simulator and the oximeter. The tests with the device also indicate that the correction factor could be quantitative for different kinds of pulse oximeters, if the calibration parameters can be achieved. The proposed correction method is simple to operate and suitable for the correction of different R value curves. It can provide certain technical support for the measurement verification of the blood oxygen module of the monitor of small brands. However, it should be noted that the attachment position, material, and thickness of the filter media will affect the correction effect. In conclusion, the proposed method of using an optical filter medium to modify the R value curve is a feasible correction method for the calibration of pulse oximeters. It has the potential to be applied to different kinds of pulse oximeters and could be of great value in the field of medical measurement.-
Key words:
- Lambert-Beer law /
- blood oxygen saturation /
- simulator /
- R value curve /
- verification
-
表 1 不同R值曲线的血氧饱和度测量值
Table 1. Oxygen saturation measurements with different R value curves of the monitor
标准值
(%)Nellcor
(%)BCI
(%)Datascope
(%)Datex
(%)Masimo
(%)100 89 99 / / / 95 89 99 / / / 90 89 99 / / / 85 89 99 / / / 80 89 99 / / / 70 89 99 / / / 注:“/”表示未显示任何数值。 表 2 加滤光介质后监护仪血氧模块测量值
Table 2. Measurement value of blood oxygen module of monitor after adding optical filter medium
标准值
(%)Nellcor
(%)BCI
(%)Datascope
(%)Datex
(%)Masimo
(%)100 100 100 100 100 100 95 95 95 94 95 95 90 90 91 89 91 92 85 84 86 83 84 85 80 79 81 78 78 80 70 72 71 65 67 69 表 3 加滤光介质后监护仪血氧模块示值重复性
Table 3. Indication repeatability of blood oxygen module of monitor after adding optical filter medium
标准值(%) R值曲线 被检监护仪测量值(%) 示值重复性(%) 100 Nellcor 100 100 100 100 100 100 0.0 90 90 90 91 90 91 91 0.5 70 71 72 71 71 72 71 0.5 100 BCI 100 100 100 100 100 100 0.0 90 91 91 90 90 91 91 0.5 70 72 71 71 72 72 72 0.5 100 Datascope 100 100 100 100 100 100 0.0 90 89 88 88 89 89 88 0.5 70 65 66 66 65 64 65 0.8 100 Datex 100 100 100 100 100 100 0.0 90 91 92 92 91 92 91 0.5 70 67 68 67 67 68 68 0.5 100 Masimo 100 100 100 100 100 100 0.0 90 92 92 93 93 93 93 0.5 70 69 69 70 69 69 71 0.8 -
[1] 徐楚天, 徐向天, 张玉华, 等. 动脉氧分压与脉搏血氧饱和度相关性研究及临床应用[J]. 中国医学装备, 2019, 16(3): 83-87. doi: 10.3969/J.ISSN.1672-8270.2019.03.023 [2] 严新忠, 杨静, 郭略. 人体血氧饱和度监测方法的研究. 医疗装备, 2005, 18(12): 1-3. [3] Jung S J, Lee Y D, Seo Y S, et al. Design of a low-power consumption wearable reflectance pulse oximetry for ubiquitous healthcare system[C]//International Conference on Control. IEEE, 2008: 526-529. [4] Frehill P, Chambers D, Rotariu C. Using Zigbee to Integrate Medical Devices[C]//2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE, 2007: 6717-6720. [5] Tavakoli M, Turicchia L, Sarpeshkar R. An ultra-low-power pulse oximeter implemented with an energy-efficient transimpedance amplifier[J]. IEEE Transactions on Biomedical Circuits & Systems, 2010, 4(1): 27-38. [6] Aoyagi T. Pulse oximetry and its simulation[J]. IEEE Tokyo Section Denshi Tokyo, 1990, 29: 184-187. [7] Aoyagi T, Fuse M, Kobayashi N, et al. Optics of pulse photometer No. 2[J]. Jpn J Med Elec Biol Eng, 2002, 39: 55. [8] Robert Finnerty. How to design a better pulse oximeter [J/OL]. ANALOG DEVICES. 2021: 1-5[2022-12-18].https://www.analog.com/cn/technical-articles/how-to-design-a-better-pulse-oximeter.html. [9] Aoyagi T. Pulse oximetry: its invention, theory, and future[J]. Journal of Anesthesia, 2003, 17: 259-266. doi: 10.1007/s00540-003-0192-6 [10] 曹云皓, 杨喜业, 刘春晨, 等. 有机/聚合物光探测器在医疗健康领域中的应用[J]. 高分子学报, 2022, 53(4): 307-330. doi: 10.11777/j.issn1000-3304.2021.21391 [11] Kock J P D, Tarassenko L. Pulse oximetry: theoretical and experimental models[J]. Medical & Biological Engineering & Computing, 1993, 31(3): 291-300. [12] 陈意, 李凯扬. 三波长血氧饱和度定标曲线的研究[J]. 北京生物医学工程, 2021, 40(5): 499-504. doi: 10.3969/j.issn.1002-3208.2021.05.009 [13] 郭萍, 孙卫新, 金捷, 等. 脉搏血氧仪定标曲线的研究[J]. 西安医科大学学报(中文版), 2000(2): 169-171. [14] 张磊. 血氧饱和度曲线的选取及性能测试过程中的重要影响[J]. 计量与测试技术, 2019, 46(10): 48-49,53. doi: 10.15988/j.cnki.1004-6941.2019.10.017 [15] M Vegfors, L G Lindberg, P Å Öberg, et al. Accuracy of pulse oximetry at various haematocrits and during haemolysis in anin vitro model[J]. Medical and Biological Engineering and Computing, 1993, 31(1): 135-141. [16] 王耀弘, 周中木, 张雯. 一种脉搏血氧仪校准装置的设计[J]. 计量与测试技术, 2020, 47(11): 39-41. doi: 10.15988/j.cnki.1004-6941.2020.11.012 [17] 蒲莉娜, 潘颂欣, 林宛华, 等. 脉搏血氧饱和度测量精度的影响因素分析[J]. 中国医疗器械信息, 2010, 16(6): 11-15,46. doi: 10.3969/j.issn.1006-6586.2010.06.004 [18] 刘文丽, 胡志雄, 杨世金, 等. 《JJG 1163-2019多参数监护仪检定规程》编制说明[J]. 中国医疗设备, 2021, 36(8): 148-151,164. doi: 10.3969/j.issn.1674-1633.2021.08.036 [19] 刘文丽, 胡志雄, 贾建革, 等. 多参数监护仪检定规程: JJG 1163-2019[S]. 北京: 国家市场监督管理总局, 2019.