Discussion on Measuring Performance of the Weighing Instrument in a Verification Facility for Water Meters Using a Weighing Method
-
摘要: 结合水表和水表检定规程相关规程,针对某厂家生产的一种水表检定装置,分析了启停质量法水表检定装置的最小用水量,并通过加装电子天平完善了该检定装置。Abstract: This paper investigates a verification facility for water meters based on related regulations for water meters and verification of water meters. The minimum water consumption of the verification facility, which adopts a start-stop weighing method, is analyzed. The verification facility was improved by adding an electronic balance.
-
表 1 Q3流量点不确定度计算汇总表
Table 1. Uncertainty budget at flow point Q3
序号 标准不确定度 不确定度来源 标准不确定度值 1 ${u_1}$ 称量重复性 28.8 g 2 ${u_2}$ 启停效应 10 g 3 ${u_3}$ 检定装置测量误差 28.8 g 4 ${u_4}$ 测温仪测温不准 17.3 g 5 ${u_5}$ 水表示值分辨力 5.8 g 6 ${u_6}$ 空气浮力 5.8 g 表 2 Q2、Q1流量点不确定度计算汇总表
Table 2. Uncertainty budget at flow points Q2 and Q1
序号 标准不确定度 不确定度来源 标准不确定度值 1 ${u_1}$ 称量重复性 14.4 g 2 ${u_2}$ 启停效应 10 g 3 ${u_3}$ 检定装置测量误差 14.4 g 4 ${u_4}$ 测温仪测温不准 1.73 g 5 ${u_5}$ 水表示值分辨力 5.8 g 6 ${u_6}$ 空气浮力 5.8 g 表 3 Q2流量点不确定度计算汇总表
Table 3. Uncertainty budget at flow point Q2
序号 标准不确定度 不确定度来源 标准不确定度值 1 ${u_1}$ 称量重复性 5.8 g 2 ${u_2}$ 启停效应 10 g 3 ${u_3}$ 检定装置测量误差 5.8 g 4 ${u_4}$ 测温仪测温不准 1.73 g 5 ${u_5}$ 水表示值分辨力 5.8 g 6 ${u_6}$ 空气浮力 5.8 g -
[1] 国家质量监督检验检疫总局.液体流量标准装置检定规程:JJG 164-2000[S].北京:中国计量出版社,2000. [2] 国家质量监督检验检疫总局.水表检定装置检定规程:JJG 1113-2015[S].北京:中国质检出版社,2015. [3] 孟涛, 李晨. JJG1113-2015《水表检定装置检定规程》解读[J]. 中国计量, 2015(10): 117-118. [4] 国家质量监督检验检疫总局.冷水水表检定规程:JJG 162-2009[S].北京:中国质检出版社,2009. [5] 国家质量监督检验检疫总局.称重传感器检定规程:JJG 669-2003[S].北京:中国计量出版社,2003. [6] 国际法制计量组织. 国际建议OIML R60- 2006非自动衡器[S]. 2006. [7] 国家技术监督局.非自动秤通用检定规程:JJG 555-1996[S].北京:中国计量出版社,1997. [8] 国家质量监督检验检疫总局.数字指示秤检定规程:JJG 539-2016[S].北京:中国质检出版社,2016. [9] 王池.流量测量不确定度分析[M].北京:中国计量出版社,2002. [10] 国家质量监督检验检疫总局.电子天平检定规程:JJG 1036-2008[S].北京:中国质检出版社,2008. [11] 詹志杰. 水表(第二版)[M]. 北京: 中国计量出版社, 2010. [12] 国家质量监督检验检疫总局.标准金属量器检定规程:JJG 259-2005[S].北京:中国计量出版社,2005.