GAO Tianheng, REN Tongxiang. Investigation into the Ablation Behavior of Different Polymer Matrices using LA-ICP-MS[J]. Metrology Science and Technology, 2023, 67(4): 63-69. doi: 10.12338/j.issn.2096-9015.2023.0105
Citation: GAO Tianheng, REN Tongxiang. Investigation into the Ablation Behavior of Different Polymer Matrices using LA-ICP-MS[J]. Metrology Science and Technology, 2023, 67(4): 63-69. doi: 10.12338/j.issn.2096-9015.2023.0105

Investigation into the Ablation Behavior of Different Polymer Matrices using LA-ICP-MS

doi: 10.12338/j.issn.2096-9015.2023.0105
  • Received Date: 2023-04-10
  • Accepted Date: 2023-05-08
  • Rev Recd Date: 2023-06-13
  • Available Online: 2023-07-03
  • Publish Date: 2023-04-18
  • Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) is a novel solid analysis technology, extensively utilized in geology, metallurgy, biology, and environmental studies among other fields. However, in LA-ICP-MS quantitative analysis of polymer samples, significant matrix effects exist, primarily attributed to differences in ablation behavior, resulting in sensitivity discrepancies and subsequently hindering accurate measurements. Currently, there is a paucity of research into the ablation behavior of such samples. This study selected four polymer samples with known concentrations of Pb, Cd, and Cr elements. The variances in ablation crater morphology under laser conditions of 193nm and 213nm wavelengths, and energy densities from 2.5 to 15 J·cm−2 were studied using extended depth of field microscopy imaging technology. Also, their effects on signal stability and sensitivity were examined. The results demonstrate superior ablation behavior of the 193nm laser compared to the 213nm laser. At least three mechanisms causing matrix effect differences were identified in the laser ablation process of polymer samples, leading to element-related sensitivity variances. They are sputtering loss, ablation mechanism loss, and flocculation-sedimentation loss. Consequently, better analysis accuracy can be achieved by improving ablation behavior through energy density control and short-wavelength laser selection. Considerations such as thermal conductivity, which influences ablation behavior, should be factored in when selecting matrix-matched standards. The study of corresponding standard materials and ablation behavior should be prioritized in this field, to enhance the accuracy of measurements and foster the development of applications in this domain.
  • loading
  • [1]
    Hattendorf B , Latkoczy C , Günther Detlef. Peer Reviewed: Laser Ablation-ICPMS[J]. Analytical Chemistry, 2003, 75(15): 341A-347A.
    [2]
    Martinez M, Baudelet M. Calibration strategies for elemental analysis of biological samples by LA-ICP-MS and LIBS – A review [J]. Anal Bioanal Chem, 2020, 412(1): 27-36.
    [3]
    Russo R E, Mao X, Liu H, et al. Laser ablation in analytical chemistry—a review [J]. Talanta, 2002, 57(3): 425-451.
    [4]
    刘勇胜, 胡兆初, 李明, 等. LA-ICP-MS在地质样品元素分析中的应用 [J]. 科学通报, 2013, 58(36): 3753-3769.
    [5]
    Austin C, Fryer F, Lear J, et al. Factors affecting internal standard selection for quantitative elemental bio-imaging of soft tissues by LA-ICP-MS [J]. J Anal At Spectrom, 2011, 26(7): 1494-1501.
    [6]
    Kuczelinis F, Petersen J H, Weis P, et al. Calibration of LA-ICP-MS via standard addition using dried picoliter droplets [J]. J Anal At Spectrom, 2020, 35(9): 1922-1931.
    [7]
    Douglas D N, O'reilly J, O'connor C, et al. Quantitation of the Fe spatial distribution in biological tissue by online double isotope dilution analysis with LA-ICP-MS: a strategy for estimating measurement uncertainty [J]. J Anal At Spectrom, 2016, 31(1): 270-279.
    [8]
    Dobrowolska J, Dehnhardt M, Matusch A, et al. Quantitative imaging of zinc, copper and lead in three distinct regions of the human brain by laser ablation inductively coupled plasma mass spectrometry [J]. Talanta, 2008, 74(4): 717-723.
    [9]
    冯流星, 王军. 同位素稀释-激光剥蚀-电感耦合等离子体质谱法测定生物组织样品中铁元素的含量 [J]. 分析化学, 2014, 42(4): 536-541.
    [10]
    王梅玲, 王海, 任丹华, 等. 铜铟镓硒薄膜元素含量的ICP-OES/ICP-MS分析 [J]. 计量科学与技术, 2022, 66(12): 11-15, 45.
    [11]
    池俏俏, 颜一军, 张娴. 激光剥蚀-电感耦合等离子体质谱分析中激光剥蚀池载气对信号响应影响的研究 [J]. 分析测试学报, 2013, 32(2): 233-238.
    [12]
    Fernández B, Claverie F, Pécheyran C, et al. Direct analysis of solid samples by fs-LA-ICP-MS [J]. TrAC, Trends Anal Chem, 2007, 26(10): 951-966.
    [13]
    刘娅聪, 王伟超, 令伟博, 等. 激光剥蚀串联电感耦合等离子体质谱在环境分析中的应用进展 [J]. 分析测试学报, 2021, 40(5): 767-776.
    [14]
    Stehrer T, Heitz J, Pedarnig J D, et al. LA-ICP-MS analysis of waste polymer materials [J]. Anal Bioanal Chem, 2010, 398(1): 415-424.
    [15]
    刘金辉, 郑令娜, 汪冰, 等. 激光剥蚀电感耦合等离子体质谱在生物样品定量分析中的研究进展 [J]. 分析科学学报, 2020, 36(3): 443-448.
    [16]
    Günther D, Heinrich C A. Comparison of the ablation behaviour of 266 nm Nd: YAG and 193 nm ArF excimer lasers for LA-ICP-MS analysis [J]. J Anal At Spectrom, 1999, 14(9): 1369-1374.
    [17]
    胡净宇, 王海舟. 高温合金痕量分析中激光剥蚀电感耦合等离子体质谱的分馏效应机理研究 [J]. 分析测试学报, 2009, 28(3): 277-282.
    [18]
    Pisonero J, Günther D. Femtosecond laser ablation inductively coupled plasma mass spectrometry: Fundamentals and capabilities for depth profiling analysis [J]. Mass Spectrom Rev, 2008, 27(6): 609-623.
    [19]
    Wu S, Xu C, Klaus S, et al. Study on Ablation Behaviors and Ablation Rates of a 193nm ArF Excimer Laser System for Selected Substrates in LA-ICP-MS Analysis [J]. Rock and Mineral Analysis, 2017, 36(5): 451-459.
    [20]
    Jenkins K, Goemann K, Belousov I, et al. Investigation of the Ablation Behaviour of Andradite-Grossular Garnets and Rutile with Implications for U-Pb Geochronology [J/OL]. [2023-04-10].https://onlinelibrary.wiley.com/doi/10.1111/ggr.12478.
    [21]
    Pham D, Tonge L, Cao J N, et al. Effects of polymer properties on laser ablation behaviour [J]. Smart Materials & Structures, 2002, 11(5): 668-674.
    [22]
    高天恒, 任同祥, 王松. 基于LA-ICP-MS的软质样品定量分析研究进展[J]. 计量科学与技术, 2023, 67(1): 29-38, 67.
    [23]
    冯流星, 马联弟, 王军. X荧光光谱法在聚合物样品中重金属元素的定量分析可行性研究 [J]. 计量学报, 2010, 31(z1): 162-165.
    [24]
    Gao T H, Ren T X, Zhou Y J, et al. The production of polymer reference materials for microanalysis with high homogeneity by a 3D printing method [J]. J Anal At Spectrom, 2023, 38(4): 893-901.
    [25]
    Nakano K, Nakamura T, Nakai I, et al. Development of the certified reference materials, plastics (JSAC 0631, JSAC 0632) for determination of hazardous metals by X-ray fluorescence analysis [J]. Bunseki Kagaku, 2007, 56(5): 363-370.
    [26]
    Simons C, Mans C, Hanning S, et al. Study on microscopic homogeneity of polymeric candidate reference materials BAM H001-BAM H010 by means of synchrotron mu-XRF and LA-ICP-MS [J]. J Anal At Spectrom, 2010, 25(1): 40-43.
    [27]
    Kempenaers L, Bings N H, Jeffries T E, et al. The use of LA-ICP-MS for the characterization of the micro-heterogeneity of heavy metals in BCR CRM 680 [J]. J Anal At Spectrom, 2001, 16(9): 1006-1011.
    [28]
    柯于球, 张路远, 柴辛娜, 等. 硫化物矿物LA-ICP-MS激光剥蚀元素信号响应 [J]. 高等学校化学学报, 2012, 33(2): 257-262.
    [29]
    Ye J, Hong Y, Li N. Experimental Study on Thrust Performance of Carbon Doped Liquid Working Substance Processed by Laser Ablation [J]. Chinese Journal of Lasers, 2017, 44(2): 0202001.
    [30]
    杨文武, 史光宇, 商琦, 等. 飞秒激光剥蚀电感耦合等离子体质谱在地球科学中的应用进展 [J]. 光谱学与光谱分析, 2017, 37(7): 2192-2198.
    [31]
    张再平, 王玉. 电感耦合等离子体质谱法测定大米粉中砷、镉含量及不确定度评估 [J]. 计量科学与技术, 2021, 65(12): 60-65.
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Figures(6)  / Tables(3)

    Article Metrics

    Article views (405) PDF downloads(28) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return