Citation: | XIE Taorui, MA Xueran, CHEN Da, LIANG Junhui, DING Xiang. Research on Calibration Methods for Anesthetic Depth Monitoring Devices[J]. Metrology Science and Technology, 2024, 68(5): 32-36. doi: 10.12338/j.issn.2096-9015.2024.0016 |
[1] |
定翔, 马雪然, 张杨, 等. 麻醉气体的红外吸收光谱等效模拟方法[J]. 计量学报, 2023, 44(10): 1626-1630. doi: 10.3969/j.issn.1000-1158.2023.10.20
|
[2] |
张亚妮, 阎文军. 麻醉意识指数在临床麻醉深度监测中的研究进展[J]. 中国临床新医学, 2021, 14(12): 1242-1246. doi: 10.3969/j.issn.1674-3806.2021.12.19
|
[3] |
邝鹰, 肖暨艳. 麻醉深度监测方法的进展[J]. 现代仪器与医疗, 2016, 22(2): 8-10. doi: 10.11876/mimt201602003
|
[4] |
Brown E N, Lydic R, Schiff N D. General Anesthesia, Sleep, and Coma[J]. New England Journal of Medicine, 2010, 363(27): 2638-2650. doi: 10.1056/NEJMra0808281
|
[5] |
Rampil I J. A Primer for EEG Signal Processing in Anesthesia[J]. Anesthesiology, 1998, 89(4): 980-1002. doi: 10.1097/00000542-199810000-00023
|
[6] |
Bruhn J, Myles P S, Sneyd R, et al. Depth of anaesthesia monitoring: what’s available, what’s validated and what’s next?[J]. British Journal of Anaesthesia, 2006, 97(1): 85-94. doi: 10.1093/bja/ael120
|
[7] |
Bowdle T A. Depth of Anesthesia Monitoring[J]. Anesthesiology Clinics of North America, 2006, 24(4): 793-822. doi: 10.1016/j.atc.2006.08.006
|
[8] |
曹桂林, 张建欣, 张传阳, 等. 监测麻醉深度对老年患者术中麻醉药用量及术后恢复的影响[J]. 国际麻醉学与复苏杂志, 2017, 38(3): 221-272. doi: 10.3760/cma.j.issn.1673-4378.2017.03.007
|
[9] |
方建新. 一种用于麻醉深度监测仪的CSI指数提取方法: ZL201610293066.3[P]. 2016-07-20.
|
[10] |
Anderson R E, Jakobsson J G. Cerebral state monitor, a new small handheld EEG monitor for determining depth of anaesthesia: a clinical comparison with the bispectral index during day-surgery[J]. European Journal of Anaesthesiology | EJA, 2006, 23(3): 208.
|
[11] |
Lobo F A, Saraiva A P, Nardiello I, et al. Electroencephalogram Monitoring in Anesthesia Practice[J]. Current Anesthesiology Reports, 2021, 11(3): 169-180. doi: 10.1007/s40140-021-00461-6
|
[12] |
Colombo M A, Napolitani M, Boly M, et al. The spectral exponent of the resting EEG indexes the presence of consciousness during unresponsiveness induced by propofol, xenon, and ketamine[J]. NeuroImage, 2019, 189: 631-644. doi: 10.1016/j.neuroimage.2019.01.024
|
[13] |
Purdon P L, Sampson A, Pavone K J, et al. Clinical Electroencephalography for Anesthesiologists Part I: Background and Basic Signatures[J]. Anesthesiology, 2015, 123(4): 937-960. doi: 10.1097/ALN.0000000000000841
|
[14] |
Sebel P S, Bowles S M, Saini V, et al. Eeg bispectrum predicts movement during thiopental/isoflurane anesthesia[J]. Journal of Clinical Monitoring, 1995, 11(2): 83-91. doi: 10.1007/BF01617729
|
[15] |
Johansen J W. Update on Bispectral Index monitoring[J]. Best Practice & Research Clinical Anaesthesiology, 2006, 20(1): 81-99.
|
[16] |
珠淮, 刘军. 一种脑电信号监测处理方法: ZL201710076293.5[P]. 2017-07-14.
|
[17] |
刘军, 周雅琪, 陈绍宾, 等. 基于样本熵与决策树的麻醉意识深度评价指数的研究[J]. 生物医学工程学杂志, 2015, 32(2): 434-439. doi: 10.7507/1001-5515.20150078
|
[18] |
Purdon P L, Pavone K J, Akeju O, et al. The Ageing Brain: Age-dependent changes in the electroencephalogram during propofol and sevoflurane general anaesthesia[J]. BJA: British Journal of Anaesthesia, 2015, 115(Suppl 1): i46-i57.
|
[19] |
周璞真, 左云霞. 脑电双频指数监测对胃肠外科全麻手术患者麻醉恢复时间影响的回顾性队列研究[J]. 中国循证医学杂志, 2018, 18(6): 565-569. doi: 10.7507/1672-2531.201803105
|
[20] |
Medtronic plc. BISTM MONITORING VALUE RANGE[M/OL].https://www.medtronic.com/content/dam/covidien/library/us/en/product/brain-monitoring/bis-monitoring-system-monitoring-value-range-information-sheet.pdf.
|
[21] |
Fu Y, Xu T, Xie K, et al. Comparative Evaluation of a New Depth of Anesthesia Index in ConView® System and the Bispectral Index during Total Intravenous Anesthesia: A Multicenter Clinical Trial[J]. BioMed Research International, 2019, 2019: 1-7. doi: 10.1155/2019/1014825.
|
[22] |
Lobo F A, Vacas S, Rossetti A O, et al. Does electroencephalographic burst suppression still play a role in the perioperative setting?[J]. Best Practice & Research Clinical Anaesthesiology, 2021, 35(2): 159-169.
|
[23] |
Hajat Z, Ahmad N, Andrzejowski J. The role and limitations of EEG-based depth of anaesthesia monitoring in theatres and intensive care[J]. Anaesthesia, 2017, 72(S1): 38-47. doi: 10.1111/anae.13739
|
[24] |
Musialowicz T, Mervaala E, Kälviäinen R, et al. Can BIS monitoring be used to assess the depth of propofol anesthesia in the treatment of refractory status epilepticus?[J]. Epilepsia, 2010, 51(8): 1580-1586. doi: 10.1111/j.1528-1167.2009.02514.x
|
[25] |
Jameson L C, Sloan T B. Using EEG to monitor anesthesia drug effects during surgery[J]. Journal of Clinical Monitoring and Computing, 2006, 20(6): 445-472. doi: 10.1007/s10877-006-9044-x
|